Clinical Accuracy QRS Detector with Automatic Parameter Adjustment in an Autonomous, Real-Time Physiologic Monitor
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F17%3APU126203" target="_blank" >RIV/00216305:26220/17:PU126203 - isvavai.cz</a>
Výsledek na webu
<a href="https://ieeexplore.ieee.org/document/8309112" target="_blank" >https://ieeexplore.ieee.org/document/8309112</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/GlobalSIP.2017.8309112" target="_blank" >10.1109/GlobalSIP.2017.8309112</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Clinical Accuracy QRS Detector with Automatic Parameter Adjustment in an Autonomous, Real-Time Physiologic Monitor
Popis výsledku v původním jazyce
This paper presents a computationally and temporal data-compact QRS complex detection algorithm useful in embedded real-time electrocardiogram (ECG) waveform analysis. The aim of the compact algorithms is to provide high sensitivity and specificity, i.e. diagnostically useful QRS waveform detection, in a continuous ambulatory monitor setting. The proposed detector uses a multi-level approach: QRS highlighting by means of a Truncated Discrete Time Stockwell Transform (TDTST), peak discrimination, and a trained Neural Network to reduce the number of false positive QRS detections. An optimization method is presented that automatically adjust the detector’s parameters to minimize the computational cost. Results demonstrate that the compact TDTST algorithm exhibits high QRS detection accuracy, an error rate of 0.31%, and remains applicable to real-time embedded physiologic ambulatory monitors.
Název v anglickém jazyce
Clinical Accuracy QRS Detector with Automatic Parameter Adjustment in an Autonomous, Real-Time Physiologic Monitor
Popis výsledku anglicky
This paper presents a computationally and temporal data-compact QRS complex detection algorithm useful in embedded real-time electrocardiogram (ECG) waveform analysis. The aim of the compact algorithms is to provide high sensitivity and specificity, i.e. diagnostically useful QRS waveform detection, in a continuous ambulatory monitor setting. The proposed detector uses a multi-level approach: QRS highlighting by means of a Truncated Discrete Time Stockwell Transform (TDTST), peak discrimination, and a trained Neural Network to reduce the number of false positive QRS detections. An optimization method is presented that automatically adjust the detector’s parameters to minimize the computational cost. Results demonstrate that the compact TDTST algorithm exhibits high QRS detection accuracy, an error rate of 0.31%, and remains applicable to real-time embedded physiologic ambulatory monitors.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
<a href="/cs/project/GAP102%2F12%2F2034" target="_blank" >GAP102/12/2034: Analýza vztahu mezi elektrickými ději a průtokem krve u srdečních komor</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
2017 IEEE Global Conference on Signal and Information Processing (GlobalSIP)
ISBN
978-1-5090-5990-4
ISSN
2376-4066
e-ISSN
—
Počet stran výsledku
5
Strana od-do
1005-1009
Název nakladatele
IEEE
Místo vydání
Montreal, QC, Kanada
Místo konání akce
Montreal, Kanada
Datum konání akce
14. 11. 2017
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000450053100199