System for gauge blocks contact?less calibration and 3d diagnostic
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F46747885%3A24210%2F14%3A%230006659" target="_blank" >RIV/46747885:24210/14:#0006659 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
System for gauge blocks contact?less calibration and 3d diagnostic
Popis výsledku v původním jazyce
In the paper is described system for gauge blocks (GBs) contact?less calibration and their 3D diagnostic in comparison to two classic calibration methods which are mentioned in the international standard EN ISO 3650 and which are commonly used for GBs calibration. In terms of the first calibration method ? the comparison one, is GB subjected to contact measurement and comparison of length is done by the reference GB. The second calibration method rests in the system when GB is wrung to the reference plane and as a GB length is presented optically measured distance between reference plane and free GB face. Such length is corrected by environment effects. At the newly designed calibration system is measurement of GB length carry out as contact?less usingonly white light and HeNe laser emission. GB length is then determinate as contact?less and with the direct sequence on the definition of length unit (one meter). Scanning of GBs faces is done by means of camera which also makes possible
Název v anglickém jazyce
System for gauge blocks contact?less calibration and 3d diagnostic
Popis výsledku anglicky
In the paper is described system for gauge blocks (GBs) contact?less calibration and their 3D diagnostic in comparison to two classic calibration methods which are mentioned in the international standard EN ISO 3650 and which are commonly used for GBs calibration. In terms of the first calibration method ? the comparison one, is GB subjected to contact measurement and comparison of length is done by the reference GB. The second calibration method rests in the system when GB is wrung to the reference plane and as a GB length is presented optically measured distance between reference plane and free GB face. Such length is corrected by environment effects. At the newly designed calibration system is measurement of GB length carry out as contact?less usingonly white light and HeNe laser emission. GB length is then determinate as contact?less and with the direct sequence on the definition of length unit (one meter). Scanning of GBs faces is done by means of camera which also makes possible
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JB - Senzory, čidla, měření a regulace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/TA03010663" target="_blank" >TA03010663: Nové systémy pro kontrolu délky koncových měrek a vyhodnocení kvality jejich povrchů</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of UNITECH´14 ? International Scientific Conference, Gabrovo
ISBN
—
ISSN
1313-230X
e-ISSN
—
Počet stran výsledku
9
Strana od-do
269-277
Název nakladatele
Technical University of Gabrovo
Místo vydání
Gabrovo ? Bulgaria
Místo konání akce
Gabrovo ? Bulgaria
Datum konání akce
1. 1. 2014
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—