Oscillometric Blood Pressure, Wrist Cuff Width and Influence on Pressure Waveforms
Popis výsledku v původním jazyce
Accuracy of blood pressure (BP) measurement with any method is affected by cuff size. Narrow cuff overestimates BP and wide cuff underestimates BP. Wrist cuff oscillometric BP monitors typically use narrow cuffs with bladders about 6 cm wide. The segmentof the radial artery compressed by the wrist cuff presents a hydraulic resistance (Rc) to blood flow during BP measurement. Narrow cuff presents smaller Rc than wide cuff. Changes in Rc alter the amplitude envelope of oscillometric waveforms. Narrow cuff allows blood flow under the cuff and into the hand sooner than wide cuff. The objective of this study was to demonstrate that narrow cuff alters the amplitude envelope in a manner that results in increased BP values when characteristic ratio algorithmis used. A specialized experimental system was used to acquire wrist oscillometric BP data from a narrow (6 cm) cuff and from a wide (10 cm) cuff. Data were acquired from 10 volunteers in sitting position. The results confirmed the object
Název v anglickém jazyce
Oscillometric Blood Pressure, Wrist Cuff Width and Influence on Pressure Waveforms
Popis výsledku anglicky
Accuracy of blood pressure (BP) measurement with any method is affected by cuff size. Narrow cuff overestimates BP and wide cuff underestimates BP. Wrist cuff oscillometric BP monitors typically use narrow cuffs with bladders about 6 cm wide. The segmentof the radial artery compressed by the wrist cuff presents a hydraulic resistance (Rc) to blood flow during BP measurement. Narrow cuff presents smaller Rc than wide cuff. Changes in Rc alter the amplitude envelope of oscillometric waveforms. Narrow cuff allows blood flow under the cuff and into the hand sooner than wide cuff. The objective of this study was to demonstrate that narrow cuff alters the amplitude envelope in a manner that results in increased BP values when characteristic ratio algorithmis used. A specialized experimental system was used to acquire wrist oscillometric BP data from a narrow (6 cm) cuff and from a wide (10 cm) cuff. Data were acquired from 10 volunteers in sitting position. The results confirmed the object
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA102%2F07%2F0147" target="_blank" >GA102/07/0147: Dynamické interakce magnetického pole s pevnými a kapalnými mechanickými systémy</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2009
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Trends in Biomedical Engineering
ISBN
978-80-227-3105-8
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
—
Název nakladatele
Slovak University of Technology
Místo vydání
Bratislava
Místo konání akce
Bratislava
Datum konání akce
18. 9. 2009
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—