Cuff width alters the amplitude envelope of wrist cuff pressure pulse waveforms
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F49777513%3A23220%2F10%3A00503277" target="_blank" >RIV/49777513:23220/10:00503277 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Cuff width alters the amplitude envelope of wrist cuff pressure pulse waveforms
Popis výsledku v původním jazyce
The accuracy of noninvasive blood pressure (BP) measurement with any method is affected by cuff width. Measurement with a too narrow cuff overestimates BP and measurement with a too wide cuff underestimates BP. Automatic wrist cuff BP monitors use permanently attached narrow cuffs with bladders about 6 cm wide. Such narrow cuffs should result in under-cuffing for wrist circumferences larger than 15 cm. The objective of this qualitative study was to show that a narrow wrist cuff results in increased BP values when a cuff pulse amplitude ratio algorithm is used. According to the algorithm used in this study, systolic pressure (SBP) corresponds to the point of 50% of maximal amplitude; for diastolic pressure (DBP) the ratio is 70%. Data were acquired from12 volunteers in the sitting position. The mean wrist circumference was 18 cm. The acquired cuff pulse data were used to compute SBP, mean pressure (MAP) and DBP. The mean values for a 6 cm cuff were SBP = 144 mmHg, MAP = 104 mmHg and DB
Název v anglickém jazyce
Cuff width alters the amplitude envelope of wrist cuff pressure pulse waveforms
Popis výsledku anglicky
The accuracy of noninvasive blood pressure (BP) measurement with any method is affected by cuff width. Measurement with a too narrow cuff overestimates BP and measurement with a too wide cuff underestimates BP. Automatic wrist cuff BP monitors use permanently attached narrow cuffs with bladders about 6 cm wide. Such narrow cuffs should result in under-cuffing for wrist circumferences larger than 15 cm. The objective of this qualitative study was to show that a narrow wrist cuff results in increased BP values when a cuff pulse amplitude ratio algorithm is used. According to the algorithm used in this study, systolic pressure (SBP) corresponds to the point of 50% of maximal amplitude; for diastolic pressure (DBP) the ratio is 70%. Data were acquired from12 volunteers in the sitting position. The mean wrist circumference was 18 cm. The acquired cuff pulse data were used to compute SBP, mean pressure (MAP) and DBP. The mean values for a 6 cm cuff were SBP = 144 mmHg, MAP = 104 mmHg and DB
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2010
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Physiological Measurement
ISSN
0967-3334
e-ISSN
—
Svazek periodika
31
Číslo periodika v rámci svazku
7
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
7
Strana od-do
—
Kód UT WoS článku
000279269900015
EID výsledku v databázi Scopus
—