Wrist Cuff Emulates Auscultation In Dual-Cuff Noninvasive Blood Pressure System
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F49777513%3A23220%2F12%3A43915656" target="_blank" >RIV/49777513:23220/12:43915656 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Wrist Cuff Emulates Auscultation In Dual-Cuff Noninvasive Blood Pressure System
Popis výsledku v původním jazyce
Manual and automatic blood pressure (BP) measurement methods have accuracy problems. This paper describes a new method that uses two cuffs to improve accuracy of BP measurement. A specially designed dual-cuff experimental system was used to acquire and process data from the two cuffs. The arm cuff provided cuff pressures (CP) and cuff arterial pulse waveforms. The arm cuff pulse pressure waveforms were used to compute mean arterial pressure (MAP), systolic pressure (SBP) and diastolic pressure (DBP) values. The SBP and DBP values were computed with characteristic ratio method. Wrist cuff waveforms were used for the emulation of auscultatory measurement of SBP and DBP. The obtained values were compared with values measured by the characteristic ratio arm cuff method.
Název v anglickém jazyce
Wrist Cuff Emulates Auscultation In Dual-Cuff Noninvasive Blood Pressure System
Popis výsledku anglicky
Manual and automatic blood pressure (BP) measurement methods have accuracy problems. This paper describes a new method that uses two cuffs to improve accuracy of BP measurement. A specially designed dual-cuff experimental system was used to acquire and process data from the two cuffs. The arm cuff provided cuff pressures (CP) and cuff arterial pulse waveforms. The arm cuff pulse pressure waveforms were used to compute mean arterial pressure (MAP), systolic pressure (SBP) and diastolic pressure (DBP) values. The SBP and DBP values were computed with characteristic ratio method. Wrist cuff waveforms were used for the emulation of auscultatory measurement of SBP and DBP. The obtained values were compared with values measured by the characteristic ratio arm cuff method.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/ED2.1.00%2F03.0094" target="_blank" >ED2.1.00/03.0094: Regionální inovační centrum elektrotechniky (RICE)</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
2012 International Conference on Applied Electronics
ISBN
978-80-261-0038-6
ISSN
1803-7232
e-ISSN
—
Počet stran výsledku
4
Strana od-do
131-134
Název nakladatele
Západočeská univerzita v Plzni
Místo vydání
Plzeň
Místo konání akce
Plzeň
Datum konání akce
5. 9. 2012
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—