Wrist cuff method determines diastolic pressure in dual cuff blood pressure system
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F49777513%3A23220%2F15%3A43925897" target="_blank" >RIV/49777513:23220/15:43925897 - isvavai.cz</a>
Výsledek na webu
<a href="http://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=7301062" target="_blank" >http://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=7301062</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Wrist cuff method determines diastolic pressure in dual cuff blood pressure system
Popis výsledku v původním jazyce
Single cuff noninvasive blood pressure (BP) measurements performed manually or automatically are subject to errors. We developed an experimental system that uses two cuffs in order to improve accuracy of noninvasive BP measurement. This study concentrated on a novel determination of diastolic BP using wrist cuff pulse waveforms and arm cuff pressures. The method was compared with the oscillometric method using the British Hypertension Society agreement grading system. The resulting grade was B- "good agreement". The wrist cuff method may prove to be better than single cuff methods because it is based on physiological principles.
Název v anglickém jazyce
Wrist cuff method determines diastolic pressure in dual cuff blood pressure system
Popis výsledku anglicky
Single cuff noninvasive blood pressure (BP) measurements performed manually or automatically are subject to errors. We developed an experimental system that uses two cuffs in order to improve accuracy of noninvasive BP measurement. This study concentrated on a novel determination of diastolic BP using wrist cuff pulse waveforms and arm cuff pressures. The method was compared with the oscillometric method using the British Hypertension Society agreement grading system. The resulting grade was B- "good agreement". The wrist cuff method may prove to be better than single cuff methods because it is based on physiological principles.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/ED2.1.00%2F03.0094" target="_blank" >ED2.1.00/03.0094: Regionální inovační centrum elektrotechniky (RICE)</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of the 20th International Conference on Applied Electronics 2015 (APPEL 2015)
ISBN
978-80-261-0385-1
ISSN
1803-7232
e-ISSN
—
Počet stran výsledku
4
Strana od-do
85-88
Název nakladatele
IEEE
Místo vydání
Piscataway
Místo konání akce
Plzeň, Česká republika
Datum konání akce
8. 9. 2015
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—