Gamma ray detection performance of newly developed MAPD‑3NM‑II photosensor with LaBr3 (Ce) crystal
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F49777513%3A23220%2F22%3A43965801" target="_blank" >RIV/49777513:23220/22:43965801 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/68407700:21670/22:00363575
Výsledek na webu
<a href="https://www.nature.com/articles/s41598-022-20006-z" target="_blank" >https://www.nature.com/articles/s41598-022-20006-z</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1038/s41598-022-20006-z" target="_blank" >10.1038/s41598-022-20006-z</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Gamma ray detection performance of newly developed MAPD‑3NM‑II photosensor with LaBr3 (Ce) crystal
Popis výsledku v původním jazyce
This paper presents the gamma-ray detection performance of the newly developed MAPD-3NM-II type SiPM sensor array (4 [Formula: see text] 4) with [Formula: see text](Ce) scintillator. The gamma-ray spectra of various sources have been measured in the energy range from 26 keV up to 1332 keV. The newly developed array based on MAPD-3NM-II sensors proved [Formula: see text] 22% enhancement in energy resolution in comparison to the former MAPD-3NM-I based array. The energy resolution of 662 keV gamma-rays measured by MAPD-3NM-II was 3.3% while clearly surpassing 4.25% resolution of MAPD-3NM-I predecessor. The enhancement is related to the high PDE of the new MAPD-3NM-II. Obtained results show that the new MAPD-3NM-II demonstrated good energy resolution and linearity in the studied energy region. The energy resolution of the new detector developed based on MAPD-3NM-II was better than all previous products of MAPD.
Název v anglickém jazyce
Gamma ray detection performance of newly developed MAPD‑3NM‑II photosensor with LaBr3 (Ce) crystal
Popis výsledku anglicky
This paper presents the gamma-ray detection performance of the newly developed MAPD-3NM-II type SiPM sensor array (4 [Formula: see text] 4) with [Formula: see text](Ce) scintillator. The gamma-ray spectra of various sources have been measured in the energy range from 26 keV up to 1332 keV. The newly developed array based on MAPD-3NM-II sensors proved [Formula: see text] 22% enhancement in energy resolution in comparison to the former MAPD-3NM-I based array. The energy resolution of 662 keV gamma-rays measured by MAPD-3NM-II was 3.3% while clearly surpassing 4.25% resolution of MAPD-3NM-I predecessor. The enhancement is related to the high PDE of the new MAPD-3NM-II. Obtained results show that the new MAPD-3NM-II demonstrated good energy resolution and linearity in the studied energy region. The energy resolution of the new detector developed based on MAPD-3NM-II was better than all previous products of MAPD.
Klasifikace
Druh
J<sub>SC</sub> - Článek v periodiku v databázi SCOPUS
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2022
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Scientific Reports
ISSN
2045-2322
e-ISSN
2045-2322
Svazek periodika
12
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
DE - Spolková republika Německo
Počet stran výsledku
8
Strana od-do
nestrankovano
Kód UT WoS článku
—
EID výsledku v databázi Scopus
2-s2.0-85138459297