Scintillation readout with MAPD array for gamma spectrometer
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F49777513%3A23220%2F20%3A43958945" target="_blank" >RIV/49777513:23220/20:43958945 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/68407700:21670/20:00346707
Výsledek na webu
<a href="https://iopscience.iop.org/article/10.1088/1748-0221/15/01/C01001" target="_blank" >https://iopscience.iop.org/article/10.1088/1748-0221/15/01/C01001</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/1748-0221/15/01/C01001" target="_blank" >10.1088/1748-0221/15/01/C01001</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Scintillation readout with MAPD array for gamma spectrometer
Popis výsledku v původním jazyce
In this study, we present the gamma-ray detection performance of LYSO, YSO(Ce) and BGO scintillators read out by a 9 ch. micropixel avalanche photodiode (MAPD) array with a high pixel density (PD) and photon detection efficiency (PDE). The array with an active area of 11,5x11,5mm2 was assembled using single MAPDs with an active area of 3,7x3,7mm2. It had a single output signal and was developed for gamma spectroscopy. Breakdown voltage measurements were carried out for each channel, as a result of which the optimal operating voltage for the array was found. The linearity range and energy resolution for each crystal were determined in the energy range from 30 to 1770 keV. The high pixel density of the array allowed to achieve good linearity in the studied energy range.
Název v anglickém jazyce
Scintillation readout with MAPD array for gamma spectrometer
Popis výsledku anglicky
In this study, we present the gamma-ray detection performance of LYSO, YSO(Ce) and BGO scintillators read out by a 9 ch. micropixel avalanche photodiode (MAPD) array with a high pixel density (PD) and photon detection efficiency (PDE). The array with an active area of 11,5x11,5mm2 was assembled using single MAPDs with an active area of 3,7x3,7mm2. It had a single output signal and was developed for gamma spectroscopy. Breakdown voltage measurements were carried out for each channel, as a result of which the optimal operating voltage for the array was found. The linearity range and energy resolution for each crystal were determined in the energy range from 30 to 1770 keV. The high pixel density of the array allowed to achieve good linearity in the studied energy range.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
<a href="/cs/project/EF16_019%2F0000766" target="_blank" >EF16_019/0000766: Inženýrské aplikace fyziky mikrosvěta</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2020
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Instrumentation
ISSN
1748-0221
e-ISSN
—
Svazek periodika
15
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
8
Strana od-do
0-7
Kód UT WoS článku
000525449100001
EID výsledku v databázi Scopus
2-s2.0-85081206474