Comparison of Modified Transiliac Internal Fixators: Mechanical Testing on Pelvic Models
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F49777513%3A23520%2F21%3A43960272" target="_blank" >RIV/49777513:23520/21:43960272 - isvavai.cz</a>
Výsledek na webu
<a href="https://scholar.google.cz/scholar?q=Comparison+of+Modified+Transiliac+Internal+Fixators:+Mechanical+Testing+on+Pelvic+Models&hl=cs&as_sdt=0&as_vis=1&oi=scholart" target="_blank" >https://scholar.google.cz/scholar?q=Comparison+of+Modified+Transiliac+Internal+Fixators:+Mechanical+Testing+on+Pelvic+Models&hl=cs&as_sdt=0&as_vis=1&oi=scholart</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Comparison of Modified Transiliac Internal Fixators: Mechanical Testing on Pelvic Models
Popis výsledku v původním jazyce
Vertically unstable transforaminal sacral fractures can be stabilized with several types of transiliac internal fixators (TIFI): the classical one (TIFI-C), the supraacetabular one (TIFI-A) and by dual application of TIFI (DTIFI). Pelvic models made of solid foam (Sawbones 1301) were used in the study. Mechanical loading tests were performed in order to asses the stiffness of the studied pelvic structures. The stiffness of the intact model was approximated as the slope of load/displacement curve. Then vertically unstable right-sided linear transforaminal fracture was created and subsequently fixed by TIFI-C, TIFI-A and DTIFI (each fixator for a separate model). The fixation techniques were compared based on the ratio between the stiffness of the treated and of the intact pelvis. Motion of the posterior pelvic structures and their deformations were measured using a photogrammetric system with four synchronous cameras. Loads applied at the base of sacrum and sacral base displacements were recorded by the testing device and used to assess the stiffness of the model structure. Obtained data were analysed by one way ANOVA test with post hoc comparison by Tukey HSD test. Mean stiffness ratio (±1SD) of pelvic structure was 0.638±0.005 for TIFI-C, 0.722±0.014 for TIFI-A and 0.720±0.008 for DTIFI. Dual transiliac internal fixation and supraacetabular fixation were superior to the classical one (p<0.0001), but DTIFI and TIFI-A stiffness ratios were statistically equivalent (p = 0.9112).
Název v anglickém jazyce
Comparison of Modified Transiliac Internal Fixators: Mechanical Testing on Pelvic Models
Popis výsledku anglicky
Vertically unstable transforaminal sacral fractures can be stabilized with several types of transiliac internal fixators (TIFI): the classical one (TIFI-C), the supraacetabular one (TIFI-A) and by dual application of TIFI (DTIFI). Pelvic models made of solid foam (Sawbones 1301) were used in the study. Mechanical loading tests were performed in order to asses the stiffness of the studied pelvic structures. The stiffness of the intact model was approximated as the slope of load/displacement curve. Then vertically unstable right-sided linear transforaminal fracture was created and subsequently fixed by TIFI-C, TIFI-A and DTIFI (each fixator for a separate model). The fixation techniques were compared based on the ratio between the stiffness of the treated and of the intact pelvis. Motion of the posterior pelvic structures and their deformations were measured using a photogrammetric system with four synchronous cameras. Loads applied at the base of sacrum and sacral base displacements were recorded by the testing device and used to assess the stiffness of the model structure. Obtained data were analysed by one way ANOVA test with post hoc comparison by Tukey HSD test. Mean stiffness ratio (±1SD) of pelvic structure was 0.638±0.005 for TIFI-C, 0.722±0.014 for TIFI-A and 0.720±0.008 for DTIFI. Dual transiliac internal fixation and supraacetabular fixation were superior to the classical one (p<0.0001), but DTIFI and TIFI-A stiffness ratios were statistically equivalent (p = 0.9112).
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
20301 - Mechanical engineering
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2021
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Acta Chirurgiae Orthopaedicae et Traumatologiae Cechoslovaca
ISSN
0001-5415
e-ISSN
—
Svazek periodika
88
Číslo periodika v rámci svazku
5
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
6
Strana od-do
333-338
Kód UT WoS článku
000798814800001