Modified Biosensor for Cholinesterase Inhibitors with Guinea Green B as the Color Indicator
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F60461373%3A22320%2F18%3A43916239" target="_blank" >RIV/60461373:22320/18:43916239 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/68407700:21460/18:00324294 RIV/70565813:_____/18:N0000004
Výsledek na webu
<a href="https://www.mdpi.com/2079-6374/8/3/81" target="_blank" >https://www.mdpi.com/2079-6374/8/3/81</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.3390/bios8030081" target="_blank" >10.3390/bios8030081</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Modified Biosensor for Cholinesterase Inhibitors with Guinea Green B as the Color Indicator
Popis výsledku v původním jazyce
Colorimetric biosensors of cholinesterase inhibitors are ideal for fast, reliable, and very simple detection of agents in air, in water, and on surfaces. This paper describes an innovation of the Czech Detehit biosensor, which is based on a biochemical enzymatic reaction visualized by using Ellman's reagent as a chromogenic indicator. The modification basically consists of a much more distinct color response of the biosensor, attained through optimization of the reaction system by using Guinea Green B as the indicator. The performance of the modified biosensor was verified on the chemical warfare agents (sarin, soman, cyclosarin, and VX) in water. The detection limits ascertained visually (with the naked eye) were about 0.001 µg/mL in water (exposure time 60 s, inhibition efficiency 25%).
Název v anglickém jazyce
Modified Biosensor for Cholinesterase Inhibitors with Guinea Green B as the Color Indicator
Popis výsledku anglicky
Colorimetric biosensors of cholinesterase inhibitors are ideal for fast, reliable, and very simple detection of agents in air, in water, and on surfaces. This paper describes an innovation of the Czech Detehit biosensor, which is based on a biochemical enzymatic reaction visualized by using Ellman's reagent as a chromogenic indicator. The modification basically consists of a much more distinct color response of the biosensor, attained through optimization of the reaction system by using Guinea Green B as the indicator. The performance of the modified biosensor was verified on the chemical warfare agents (sarin, soman, cyclosarin, and VX) in water. The detection limits ascertained visually (with the naked eye) were about 0.001 µg/mL in water (exposure time 60 s, inhibition efficiency 25%).
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10401 - Organic chemistry
Návaznosti výsledku
Projekt
<a href="/cs/project/VI20152018024" target="_blank" >VI20152018024: NOVÉ TECHNOLOGIE PRO OSOBNÍ DETEKTOR INHIBITORŮ CHOLINESTERÁZ</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2018
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Biosensors
ISSN
2079-6374
e-ISSN
—
Svazek periodika
8
Číslo periodika v rámci svazku
3
Stát vydavatele periodika
CH - Švýcarská konfederace
Počet stran výsledku
10
Strana od-do
—
Kód UT WoS článku
000448396500025
EID výsledku v databázi Scopus
2-s2.0-85054766054