Referentní elektrody založené na pevných amalgamech
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388955%3A_____%2F04%3A00101042" target="_blank" >RIV/61388955:_____/04:00101042 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/15273440:_____/04:00009810
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Reference Electrodes Based on Solid Amalgams
Popis výsledku v původním jazyce
A reference saturated calomel electrode based on the non-toxic silver solid amalgam (SCE-AgSA) as a substitute for liquid mercury is described. Long-term as well as short-term tests made during a period of one year confirmed the SCE-AgSA potential to beequal to that of SCE within the limits of +/-1 mV (the difference DeltaE(SCE-AgSA) = + 1.08 +/- 0.24 mV (vs. SCE); SD = 0.87; N = 5 1). The voltammograms of four selected metal ions registered vs. SCE-AgSA and vs. SCE could practically overlap each other. None of the mentioned electrodes were polarized to a significant degree even if the electric current applied reached 2 mA. On the basis of the obtained results the SCE-AgSA could be considered as equivalent substitute for SCE. Owing to practically identical electrode potential of both types of the mentioned electrodes the results obtained using SCF-AgSA are completely comparable to those measured vs. SCE, without any correction of peak potentials.
Název v anglickém jazyce
Reference Electrodes Based on Solid Amalgams
Popis výsledku anglicky
A reference saturated calomel electrode based on the non-toxic silver solid amalgam (SCE-AgSA) as a substitute for liquid mercury is described. Long-term as well as short-term tests made during a period of one year confirmed the SCE-AgSA potential to beequal to that of SCE within the limits of +/-1 mV (the difference DeltaE(SCE-AgSA) = + 1.08 +/- 0.24 mV (vs. SCE); SD = 0.87; N = 5 1). The voltammograms of four selected metal ions registered vs. SCE-AgSA and vs. SCE could practically overlap each other. None of the mentioned electrodes were polarized to a significant degree even if the electric current applied reached 2 mA. On the basis of the obtained results the SCE-AgSA could be considered as equivalent substitute for SCE. Owing to practically identical electrode potential of both types of the mentioned electrodes the results obtained using SCF-AgSA are completely comparable to those measured vs. SCE, without any correction of peak potentials.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CG - Elektrochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/OE92" target="_blank" >OE92: Kombinovaný chronopotenciometrický/voltametrický mobilní (přenosný) analyzátor pro určování toxických a dalších látek ve vodných roztocích a v biologických tekutinách.</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2004
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Electroanalysis
ISSN
1040-0397
e-ISSN
—
Svazek periodika
16
Číslo periodika v rámci svazku
3
Stát vydavatele periodika
DE - Spolková republika Německo
Počet stran výsledku
4
Strana od-do
238-241
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—