Obecná metoda výpočtu koncentrací stopových látek ve vzduchu a dechu z dat hmotnostní spektrometrie v proudové trubici s vybranými ionty
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388955%3A_____%2F06%3A00078415" target="_blank" >RIV/61388955:_____/06:00078415 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
A general method for the calculation of absolute trace gas concentrations in air and breath from selected ion flow tube mass spectrometry data
Popis výsledku v původním jazyce
A complete description is presented of a numerical method that allows the calculation, in real time, of absolute concentrations of trace gases, including volatile organic compounds and water vapour, from selected ion flow tube mass spectrometry, SIFT-MS,data. No assumptions are made concerning the SIFT-MS instrument size or its configuration and thus the calculation can be applied to the currently available, relatively large instruments and the anticipated new generation of smaller SIFT-MS instruments.This numerical method clearly distinguishes those parameters that are obviously specific to a particular instrument, including flow tube geometry, degree of mass discrimination in the analytical mass spectrometer and flow tube reaction time, from general fundamental processes, in particular the differential diffusive loss of ions along the flow tube that is dependent on the properties of those ions involved in the determination of the concentrations of particular trace gases.
Název v anglickém jazyce
A general method for the calculation of absolute trace gas concentrations in air and breath from selected ion flow tube mass spectrometry data
Popis výsledku anglicky
A complete description is presented of a numerical method that allows the calculation, in real time, of absolute concentrations of trace gases, including volatile organic compounds and water vapour, from selected ion flow tube mass spectrometry, SIFT-MS,data. No assumptions are made concerning the SIFT-MS instrument size or its configuration and thus the calculation can be applied to the currently available, relatively large instruments and the anticipated new generation of smaller SIFT-MS instruments.This numerical method clearly distinguishes those parameters that are obviously specific to a particular instrument, including flow tube geometry, degree of mass discrimination in the analytical mass spectrometer and flow tube reaction time, from general fundamental processes, in particular the differential diffusive loss of ions along the flow tube that is dependent on the properties of those ions involved in the determination of the concentrations of particular trace gases.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CF - Fyzikální chemie a teoretická chemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA202%2F03%2F0827" target="_blank" >GA202/03/0827: Studium elementárních procesů v nízkoteplotním a technologicky orientovaném plazmatu a vývoj relevantních diagnostických metod</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2006
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
International Journal of Mass Spectrometry
ISSN
1387-3806
e-ISSN
—
Svazek periodika
249-250
Číslo periodika v rámci svazku
-
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
10
Strana od-do
230-239
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—