Stress analysis of the hip joint endoprosthesis ceramic head for different values of shape deviations
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388998%3A_____%2F14%3A00441585" target="_blank" >RIV/61388998:_____/14:00441585 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.4028/www.scientific.net/AMR.936.770" target="_blank" >http://dx.doi.org/10.4028/www.scientific.net/AMR.936.770</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.4028/www.scientific.net/AMR.936.770" target="_blank" >10.4028/www.scientific.net/AMR.936.770</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Stress analysis of the hip joint endoprosthesis ceramic head for different values of shape deviations
Popis výsledku v původním jazyce
The paper deals with the problems of ceramic head of hip joint endoprosthesis destructions, and with assessing the impact of shape deviations of conical surfaces on the tensile stress under ISO 7206-5 loading. The failure of the hip joint endoprosthesisceramic head has always traumatic consequences for the patient, since a part of or even the whole endoprosthesis has to be reoperated. Hence, it is desired to reduce the number of implant re-operations to the minimum. Therefore the computational modelling of the stress of the head was realised. The shape deviations of the ideal contact cone areas of the head and stem are parameters that significantly influence the tensile stress in the head and its reliability. The assumed shape deviations of the head'sand stem's cones are macro shape deviations (different cone taper) and micro shape deviations (unevennes) measured using the IMS-UMPIRE equipment. The stress state in the ceramic heads was solved using the FEM and head's failure probabil
Název v anglickém jazyce
Stress analysis of the hip joint endoprosthesis ceramic head for different values of shape deviations
Popis výsledku anglicky
The paper deals with the problems of ceramic head of hip joint endoprosthesis destructions, and with assessing the impact of shape deviations of conical surfaces on the tensile stress under ISO 7206-5 loading. The failure of the hip joint endoprosthesisceramic head has always traumatic consequences for the patient, since a part of or even the whole endoprosthesis has to be reoperated. Hence, it is desired to reduce the number of implant re-operations to the minimum. Therefore the computational modelling of the stress of the head was realised. The shape deviations of the ideal contact cone areas of the head and stem are parameters that significantly influence the tensile stress in the head and its reliability. The assumed shape deviations of the head'sand stem's cones are macro shape deviations (different cone taper) and micro shape deviations (unevennes) measured using the IMS-UMPIRE equipment. The stress state in the ceramic heads was solved using the FEM and head's failure probabil
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JH - Keramika, žáruvzdorné materiály a skla
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA13-34632S" target="_blank" >GA13-34632S: Zvýšení úrovně výpočtového modelování chování keramické hlavice totální kyčelní endoprotézy</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Advanced Materials Research
ISBN
978-3-03835-102-3
ISSN
1022-6680
e-ISSN
—
Počet stran výsledku
6
Strana od-do
770-775
Název nakladatele
Trans Tech Publications Ltd
Místo vydání
Dürnten
Místo konání akce
Shanghai
Datum konání akce
28. 6. 2014
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—