Tensile stress analyses of the hip joint endoprosthesis ceramic head with real shape deviations
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388998%3A_____%2F15%3A00441584" target="_blank" >RIV/61388998:_____/15:00441584 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1007/978-3-319-11776-8_111" target="_blank" >http://dx.doi.org/10.1007/978-3-319-11776-8_111</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1007/978-3-319-11776-8_111" target="_blank" >10.1007/978-3-319-11776-8_111</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Tensile stress analyses of the hip joint endoprosthesis ceramic head with real shape deviations
Popis výsledku v původním jazyce
In vivo fracture of ceramic heads of total hip joint endoprostheses has been stated in a not negligible number of patients in the Czech Republic hospitals, some time ago. The reliability of the ceramic head is based on the Weibull weakest link theory. The probability of the failure depends on the tensile stress values in the head, on the volume in which stress acts and on the material parameters of the used bioceramics. The stress analyses of the ceramic heads were realized using computational modelling-FEM system ANSYS under ISO 7206-5 loading. The maximum values of the tensile stress in the ceramic head are significantly influenced on the shape deviations of conical surfaces of the head and the stem cones. There are two types of the shape deviations–the first is macro deviations (different angle of the taper of the stem cone and the head cone), the second is micro shape deviations (the stochastic distribution of unevenness of cone contact surfaces in order micrometers). The macro and micro shape deviations of the real cones were carried out using the IMS-UMPIRE measuring equipment. By computational modelling it has been proved that the character and the size of shape deviations on a conical contact area between the stem and the head of hip joint endoprosthesis have a pronounced influence on the character and the value of stress in the head and, hence, on the probability of the head's failure. Ignorance of this influence can lead to unforeseen failures of ceramic heads and thus to cast doubt on the trustworthiness of a certain type of total hip joint endoprosthesis and of surgical treatment.
Název v anglickém jazyce
Tensile stress analyses of the hip joint endoprosthesis ceramic head with real shape deviations
Popis výsledku anglicky
In vivo fracture of ceramic heads of total hip joint endoprostheses has been stated in a not negligible number of patients in the Czech Republic hospitals, some time ago. The reliability of the ceramic head is based on the Weibull weakest link theory. The probability of the failure depends on the tensile stress values in the head, on the volume in which stress acts and on the material parameters of the used bioceramics. The stress analyses of the ceramic heads were realized using computational modelling-FEM system ANSYS under ISO 7206-5 loading. The maximum values of the tensile stress in the ceramic head are significantly influenced on the shape deviations of conical surfaces of the head and the stem cones. There are two types of the shape deviations–the first is macro deviations (different angle of the taper of the stem cone and the head cone), the second is micro shape deviations (the stochastic distribution of unevenness of cone contact surfaces in order micrometers). The macro and micro shape deviations of the real cones were carried out using the IMS-UMPIRE measuring equipment. By computational modelling it has been proved that the character and the size of shape deviations on a conical contact area between the stem and the head of hip joint endoprosthesis have a pronounced influence on the character and the value of stress in the head and, hence, on the probability of the head's failure. Ignorance of this influence can lead to unforeseen failures of ceramic heads and thus to cast doubt on the trustworthiness of a certain type of total hip joint endoprosthesis and of surgical treatment.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JH - Keramika, žáruvzdorné materiály a skla
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA13-34632S" target="_blank" >GA13-34632S: Zvýšení úrovně výpočtového modelování chování keramické hlavice totální kyčelní endoprotézy</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
5th International Conference on Biomedical Engineering in Vietnam
ISBN
978-3-319-11775-1
ISSN
1680-0737
e-ISSN
—
Počet stran výsledku
4
Strana od-do
450-453
Název nakladatele
Springer Verlag
Místo vydání
NEW YORK
Místo konání akce
Ho Chi Minh City
Datum konání akce
16. 6. 2014
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000357778100111