Napjatostní a spolehlivostní analýza keramické hlavice totální kyc(elní endoprotézy s makro a mikro tvarovými odchylkami
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26210%2F08%3APU79937" target="_blank" >RIV/00216305:26210/08:PU79937 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/61388998:_____/08:00322478
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Stress and Reliability Analyses of the Hip Joint Endoprosthesis Ceramic Head with Macro and Micro Shape Deviations
Popis výsledku v původním jazyce
The principal aim of this work is finding the reasons of the head's fractures because in the Czech Republic, some time ago, in vivo fracture of ceramic heads of total hip joint endoprostheses has been stated in a not negligible number of patients. The reliability of the ceramic head is based on the Weibull weakest link theory. The probability of the failure depends on the tensile stress values in the head, on the volume in which stress acts and on the material parameters of the used bioceramics (aluminium oxide). The stress analyses of the ceramic heads were realized using computational modelling - FEM system ANSYS under ISO 7206-5 loading. The maximum values of the tensile stress (the first principal stress) in the ceramic head are significantly influenced on the shape deviations of conical surfaces of the head and the stem cones. There are two types of the shape deviations - the first is macro deviations (different angle of the taper of the stem cone and the head cone), the second is
Název v anglickém jazyce
Stress and Reliability Analyses of the Hip Joint Endoprosthesis Ceramic Head with Macro and Micro Shape Deviations
Popis výsledku anglicky
The principal aim of this work is finding the reasons of the head's fractures because in the Czech Republic, some time ago, in vivo fracture of ceramic heads of total hip joint endoprostheses has been stated in a not negligible number of patients. The reliability of the ceramic head is based on the Weibull weakest link theory. The probability of the failure depends on the tensile stress values in the head, on the volume in which stress acts and on the material parameters of the used bioceramics (aluminium oxide). The stress analyses of the ceramic heads were realized using computational modelling - FEM system ANSYS under ISO 7206-5 loading. The maximum values of the tensile stress (the first principal stress) in the ceramic head are significantly influenced on the shape deviations of conical surfaces of the head and the stem cones. There are two types of the shape deviations - the first is macro deviations (different angle of the taper of the stem cone and the head cone), the second is
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BO - Biofyzika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/1P05ME789" target="_blank" >1P05ME789: Simulace mechanické funkce vybraných prvků lidského těla</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2008
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
13th International Conference on Biomedical Engineering
ISBN
978-3-540-92840-9
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
—
Název nakladatele
Springer
Místo vydání
Singapore
Místo konání akce
Singapore
Datum konání akce
3. 12. 2008
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—