Calibration of modified Liulin detector for cosmic radiation measurements on-board aircraft
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61389005%3A_____%2F15%3A00445293" target="_blank" >RIV/61389005:_____/15:00445293 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1093/rpd/ncv332" target="_blank" >http://dx.doi.org/10.1093/rpd/ncv332</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/rpd/ncv332" target="_blank" >10.1093/rpd/ncv332</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Calibration of modified Liulin detector for cosmic radiation measurements on-board aircraft
Popis výsledku v původním jazyce
The annual effective doses of aircrew members often exceed the limit of 1 mSv for the public due to the increased level of cosmic radiation at the flight altitudes, and thus, it is recommended to monitor them. Aircrew dosimetry is usually performed usingspecial computer programs mostly based on results of Monte Carlo simulations. Contemporary, detectors are used mostly for validation of these computer codes, verification of effective dose calculations and for research purposes. One of such detectors isactive silicon semiconductor deposited energy spectrometer Liulin. Output quantities of measurement with the Liulin detector are the absorbed dose in silicon D and the ambient dose equivalent H*(10); to determine it, two calibrations are necessary. Thepurpose of this work was to develop a calibration methodology that can be used to convert signal from the detector to D independently on calibration performed at Heavy Ion Medical Accelerator facility in Chiba, Japan.
Název v anglickém jazyce
Calibration of modified Liulin detector for cosmic radiation measurements on-board aircraft
Popis výsledku anglicky
The annual effective doses of aircrew members often exceed the limit of 1 mSv for the public due to the increased level of cosmic radiation at the flight altitudes, and thus, it is recommended to monitor them. Aircrew dosimetry is usually performed usingspecial computer programs mostly based on results of Monte Carlo simulations. Contemporary, detectors are used mostly for validation of these computer codes, verification of effective dose calculations and for research purposes. One of such detectors isactive silicon semiconductor deposited energy spectrometer Liulin. Output quantities of measurement with the Liulin detector are the absorbed dose in silicon D and the ambient dose equivalent H*(10); to determine it, two calibrations are necessary. Thepurpose of this work was to develop a calibration methodology that can be used to convert signal from the detector to D independently on calibration performed at Heavy Ion Medical Accelerator facility in Chiba, Japan.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BG - Jaderná, atomová a molekulová fyzika, urychlovače
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/LG13031" target="_blank" >LG13031: Spolupráce ČR s CERN</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Radiation Protection Dosimetry
ISBN
—
ISSN
1742-3406
e-ISSN
—
Počet stran výsledku
4
Strana od-do
1-4
Název nakladatele
Oxford Journals
Místo vydání
Oxford
Místo konání akce
Prague
Datum konání akce
1. 9. 2014
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—