CALIBRATION OF SILICON DETECTORS LIULIN AND AIRDOS USING COSMIC RAYS AND TIMEPIX FOR USE AT FLIGHT ALTITUDES
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21340%2F22%3A00359667" target="_blank" >RIV/68407700:21340/22:00359667 - isvavai.cz</a>
Výsledek na webu
<a href="https://doi.org/10.1093/rpd/ncac104" target="_blank" >https://doi.org/10.1093/rpd/ncac104</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/rpd/ncac104" target="_blank" >10.1093/rpd/ncac104</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
CALIBRATION OF SILICON DETECTORS LIULIN AND AIRDOS USING COSMIC RAYS AND TIMEPIX FOR USE AT FLIGHT ALTITUDES
Popis výsledku v původním jazyce
Silicon detectors such as Liulin and AIRDOS are used for cosmic radiation measurements onboard aircraft. These measurements can be used for the verification of computer programs assessing aircraft crew radiation exposure. Recently performed intercomparison flights showed large variances of absorbed doses among individual detectors and significant differences between results of silicon detectors and computer programs. In order to explain for these differences, we have developed energy calibration method that can be performed on short flights. The method is based on cross-calibration of Liulin and AIRDOS deposited energy spectra with deposited energy spectra measured by Timepix which has superior detection properties in terms of energy resolution and the detection threshold. Moreover, the portion of dose which is omitted due to low sensitivity for low-energy deposits was calculated. The resulting absorbed dose rates at two intercomparison flights show significantly improved variation of results and better agreement with modelled absorbed dose rates.
Název v anglickém jazyce
CALIBRATION OF SILICON DETECTORS LIULIN AND AIRDOS USING COSMIC RAYS AND TIMEPIX FOR USE AT FLIGHT ALTITUDES
Popis výsledku anglicky
Silicon detectors such as Liulin and AIRDOS are used for cosmic radiation measurements onboard aircraft. These measurements can be used for the verification of computer programs assessing aircraft crew radiation exposure. Recently performed intercomparison flights showed large variances of absorbed doses among individual detectors and significant differences between results of silicon detectors and computer programs. In order to explain for these differences, we have developed energy calibration method that can be performed on short flights. The method is based on cross-calibration of Liulin and AIRDOS deposited energy spectra with deposited energy spectra measured by Timepix which has superior detection properties in terms of energy resolution and the detection threshold. Moreover, the portion of dose which is omitted due to low sensitivity for low-energy deposits was calculated. The resulting absorbed dose rates at two intercomparison flights show significantly improved variation of results and better agreement with modelled absorbed dose rates.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10301 - Atomic, molecular and chemical physics (physics of atoms and molecules including collision, interaction with radiation, magnetic resonances, Mössbauer effect)
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2022
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Protection Dosimetry
ISSN
0144-8420
e-ISSN
1742-3406
Svazek periodika
198
Číslo periodika v rámci svazku
9-11
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
7
Strana od-do
597-603
Kód UT WoS článku
000844405700018
EID výsledku v databázi Scopus
—