Relative efficiency of radiophotoluminescent glass dosimeters in a scanning pencil proton beam
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61389005%3A_____%2F24%3A00581046" target="_blank" >RIV/61389005:_____/24:00581046 - isvavai.cz</a>
Výsledek na webu
<a href="https://doi.org/10.1016/j.radphyschem.2023.111396" target="_blank" >https://doi.org/10.1016/j.radphyschem.2023.111396</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.radphyschem.2023.111396" target="_blank" >10.1016/j.radphyschem.2023.111396</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Relative efficiency of radiophotoluminescent glass dosimeters in a scanning pencil proton beam
Popis výsledku v původním jazyce
The increasing role of hadron therapy, and particularly of proton therapy, in cancer treatment requires dosimetric characterization of luminescent detectors in heavy charged particle beams. The ionization density, which is related to the linear energy transfer (LET), is changing when heavy charged particle is penetrating through a material. Therefore, it is crucial to evaluate how efficiency of the used dosimeters depends on the change in LET.The aim of this study was to evaluate relative efficiency of radiophotoluminescent (RPL) glass dosimeters (type GD-352M) in a clinical scanning pencil proton beam of 100 MeV and compare it to the results obtained with thermoluminescent (TL) dosimeters (type TLD-100). LETf and LETd at the selected dosimeter's position were evaluated and compared by two Monte Carlo codes (MCNP and PHITS). Linear dose response is precondition for the evaluation of dosimeter efficiency and first, dose response in proton beam was evaluated and compared with those in 60Co gamma ray field.Linear dose response up to 9 Gy was confirmed for RPLs in proton beam, and gamma ray field. For TLDs, supralinearity from about 5 Gy was observed. A constant relative efficiency is observed along the plateau (corresponding to LETf from 0.7 keV/mu m to 1.7 keV/mu m) of non-modulated proton beam with an average value 1.00 +/- 0.02 for RPL and 0.97 +/- 0.02 for TL dosimeters. Relative efficiency decrease of 10 % is observed for RPLs within the Bragg peak with LETf = 2.3 keV/mu m. Future work will be focused on the evaluation of the efficiency for higher LET values using more energetic proton beams and precise positioning along the Bragg peak.
Název v anglickém jazyce
Relative efficiency of radiophotoluminescent glass dosimeters in a scanning pencil proton beam
Popis výsledku anglicky
The increasing role of hadron therapy, and particularly of proton therapy, in cancer treatment requires dosimetric characterization of luminescent detectors in heavy charged particle beams. The ionization density, which is related to the linear energy transfer (LET), is changing when heavy charged particle is penetrating through a material. Therefore, it is crucial to evaluate how efficiency of the used dosimeters depends on the change in LET.The aim of this study was to evaluate relative efficiency of radiophotoluminescent (RPL) glass dosimeters (type GD-352M) in a clinical scanning pencil proton beam of 100 MeV and compare it to the results obtained with thermoluminescent (TL) dosimeters (type TLD-100). LETf and LETd at the selected dosimeter's position were evaluated and compared by two Monte Carlo codes (MCNP and PHITS). Linear dose response is precondition for the evaluation of dosimeter efficiency and first, dose response in proton beam was evaluated and compared with those in 60Co gamma ray field.Linear dose response up to 9 Gy was confirmed for RPLs in proton beam, and gamma ray field. For TLDs, supralinearity from about 5 Gy was observed. A constant relative efficiency is observed along the plateau (corresponding to LETf from 0.7 keV/mu m to 1.7 keV/mu m) of non-modulated proton beam with an average value 1.00 +/- 0.02 for RPL and 0.97 +/- 0.02 for TL dosimeters. Relative efficiency decrease of 10 % is observed for RPLs within the Bragg peak with LETf = 2.3 keV/mu m. Future work will be focused on the evaluation of the efficiency for higher LET values using more energetic proton beams and precise positioning along the Bragg peak.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
30224 - Radiology, nuclear medicine and medical imaging
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2024
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Physics and Chemistry
ISSN
0969-806X
e-ISSN
1879-0895
Svazek periodika
216
Číslo periodika v rámci svazku
MAR
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
9
Strana od-do
111396
Kód UT WoS článku
001132601400001
EID výsledku v databázi Scopus
2-s2.0-85178160954