Srovnání terapeutického monitorování cyklosporinu A a jeho metabolitů u pacientů s transplantovanými ledvinami při použití HPLC a RIA metod
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61988987%3A17110%2F01%3AA0900ANI" target="_blank" >RIV/61988987:17110/01:A0900ANI - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Evaluation and comparison of therapeutic monitoring of whole-blood levels of cyclosporin A and its metabolites in renal transplantation by HPLC and RIA methods
Popis výsledku v původním jazyce
Background: The aim of the work was to evaluate the possibility to estimate the level of cyclosporin A (CyA) metabolites as the difference of radioimmunoassay (RIA) non-specific and RIA specific methods. Methods: Blood samples of renal transplant patients were analyzed by three different methods: RIA specific method (CYCLO-Trac, DiaSorin, USA) (RIA(SP)), RIA non-specific method (Immunotech, Czech Republic) (RIA(NS)), and high performance liquid chromatography (HPLC) method. Conclusion: The combination of both the specific and non-specific methods for the determination of CyA presents an improved means for the TDM of CyA and CyA metabolites in renal transplant patients. Moreover, a combination of both methods can help to elucidate some unexpected events, such as the persistence of high cyclosporin blood levels. (C) 2001 Elsevier Science B.V. All rights reserved.
Název v anglickém jazyce
Evaluation and comparison of therapeutic monitoring of whole-blood levels of cyclosporin A and its metabolites in renal transplantation by HPLC and RIA methods
Popis výsledku anglicky
Background: The aim of the work was to evaluate the possibility to estimate the level of cyclosporin A (CyA) metabolites as the difference of radioimmunoassay (RIA) non-specific and RIA specific methods. Methods: Blood samples of renal transplant patients were analyzed by three different methods: RIA specific method (CYCLO-Trac, DiaSorin, USA) (RIA(SP)), RIA non-specific method (Immunotech, Czech Republic) (RIA(NS)), and high performance liquid chromatography (HPLC) method. Conclusion: The combination of both the specific and non-specific methods for the determination of CyA presents an improved means for the TDM of CyA and CyA metabolites in renal transplant patients. Moreover, a combination of both methods can help to elucidate some unexpected events, such as the persistence of high cyclosporin blood levels. (C) 2001 Elsevier Science B.V. All rights reserved.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FR - Farmakologie a lékárnická chemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
R - Projekt Ramcoveho programu EK
Ostatní
Rok uplatnění
2001
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Clin. Chim. Acta
ISSN
0009-8981
e-ISSN
—
Svazek periodika
310
Číslo periodika v rámci svazku
20
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
6
Strana od-do
—
Kód UT WoS článku
000170811100008
EID výsledku v databázi Scopus
—