STATISTICAL MODEL OF QUALITY OF RADON MEASUREMENTS USING ELECTRET ION CHAMBER DETECTORS
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27240%2F11%3A86080828" target="_blank" >RIV/61989100:27240/11:86080828 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/61989100:27200/11:86080828
Výsledek na webu
<a href="http://dx.doi.org/10.1093/rpd/ncr058" target="_blank" >http://dx.doi.org/10.1093/rpd/ncr058</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1093/rpd/ncr058" target="_blank" >10.1093/rpd/ncr058</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
STATISTICAL MODEL OF QUALITY OF RADON MEASUREMENTS USING ELECTRET ION CHAMBER DETECTORS
Popis výsledku v původním jazyce
This paper presents a statistical model for estimating probability, which states that a quality of radon measurements using the electret ion chamber system RM-1 will be accurate. The quality of the electret measurement was modelled as a ratio between thereference radon concentration and values measured under real conditions with varied levels of indoor radon concentration and microclimate indoor conditions. It was stated that the accuracy tolerance of measured values is 20 % of the given reference value. To estimate the uncertainty of the statistical model, the exact confidence limits for the estimated probabilities are computed. The statistical model was confirmed by an independent set of measurements. Moreover, the effect of absolute humidity on thequality estimation of electret detectors are also statistically analysed and discussed. The results of the statistical model confirm that the electret system is robust and suitable for estimation of radon concentration.
Název v anglickém jazyce
STATISTICAL MODEL OF QUALITY OF RADON MEASUREMENTS USING ELECTRET ION CHAMBER DETECTORS
Popis výsledku anglicky
This paper presents a statistical model for estimating probability, which states that a quality of radon measurements using the electret ion chamber system RM-1 will be accurate. The quality of the electret measurement was modelled as a ratio between thereference radon concentration and values measured under real conditions with varied levels of indoor radon concentration and microclimate indoor conditions. It was stated that the accuracy tolerance of measured values is 20 % of the given reference value. To estimate the uncertainty of the statistical model, the exact confidence limits for the estimated probabilities are computed. The statistical model was confirmed by an independent set of measurements. Moreover, the effect of absolute humidity on thequality estimation of electret detectors are also statistically analysed and discussed. The results of the statistical model confirm that the electret system is robust and suitable for estimation of radon concentration.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BB - Aplikovaná statistika, operační výzkum
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Protection Dosimetry
ISSN
0144-8420
e-ISSN
—
Svazek periodika
145
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
5
Strana od-do
295-299
Kód UT WoS článku
000290818900040
EID výsledku v databázi Scopus
—