Refining the diagnostic quality of the abdominal fetal electrocardiogram using the techniques of artificial intelligence.
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989100%3A27240%2F12%3A86083669" target="_blank" >RIV/61989100:27240/12:86083669 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Refining the diagnostic quality of the abdominal fetal electrocardiogram using the techniques of artificial intelligence.
Popis výsledku v původním jazyce
This article deals with utilization of the combination of the fuzzy system and artificial intelligence techniques, called the Adaptive Neuro Fuzzy Inference System ANFIS, with the aim to refine the diagnostic quality of the abdominal fetal electrocardiogram FECG. Within the scope of the experiments carried out and based on the ANFIS structure the authors created a complex system for removing the undesirable mother?s MECG degrading the abdominal FECG. Current research shows that the application of the conventional systems for enhancing the diagnostic quality of the abdominal FECG faces a series of problems (e.g. non-linear character of the task to solve, computational complexity of RLS algorithms, etc.). The need for a higher diagnostic quality of the abdominal FECG is reflected in the authors? intention to utilize the designed system for the latest intrapartum monitoring method, called ST analysis. In terms of this advanced method, the aspect subjected to a diagnostic analysis is the S
Název v anglickém jazyce
Refining the diagnostic quality of the abdominal fetal electrocardiogram using the techniques of artificial intelligence.
Popis výsledku anglicky
This article deals with utilization of the combination of the fuzzy system and artificial intelligence techniques, called the Adaptive Neuro Fuzzy Inference System ANFIS, with the aim to refine the diagnostic quality of the abdominal fetal electrocardiogram FECG. Within the scope of the experiments carried out and based on the ANFIS structure the authors created a complex system for removing the undesirable mother?s MECG degrading the abdominal FECG. Current research shows that the application of the conventional systems for enhancing the diagnostic quality of the abdominal FECG faces a series of problems (e.g. non-linear character of the task to solve, computational complexity of RLS algorithms, etc.). The need for a higher diagnostic quality of the abdominal FECG is reflected in the authors? intention to utilize the designed system for the latest intrapartum monitoring method, called ST analysis. In terms of this advanced method, the aspect subjected to a diagnostic analysis is the S
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/LH12183" target="_blank" >LH12183: Smart instrumentation - Vývoj a ověřování pokročilých měřicích systémů pro aplikační oblast přenosu a distribuce elektrické energie</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Przeglad elektrotechniczny
ISSN
0033-2097
e-ISSN
—
Svazek periodika
R. 88
Číslo periodika v rámci svazku
12b/2012
Stát vydavatele periodika
PL - Polská republika
Počet stran výsledku
6
Strana od-do
155-160
Kód UT WoS článku
000314689600046
EID výsledku v databázi Scopus
—