Wavense technology glucometer Linus for routine self-monitoring and clinical practice
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989592%3A15110%2F11%3A10213183" target="_blank" >RIV/61989592:15110/11:10213183 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Wavense technology glucometer Linus for routine self-monitoring and clinical practice
Popis výsledku v původním jazyce
Conventional glycometer systems for plasma/blood glucose monitoring are based colorimetry or static electrochemistry using a fixed input signal. The recent glucometer Linus, Wellion, Agamatrix, USA, based on wavesense dynamic electrochemistry, uses a time-varying input signal to give a more accurate glucose reading. The purpose of this study was to compare the plasma glucose (PG) readings obtained by nursing staff from glucometer Linus and PG values estimated on an approved analyzer Daytona TM, Randox,Global Medical Instrumentation, Inc., MN, USA. Glucometer Linus readings correspond to the ISO and FDA standards and meet the criteria of clinical utility. So, Linus appears to be an accurate and reliable device for PG-self-monitoring and clinical practice.
Název v anglickém jazyce
Wavense technology glucometer Linus for routine self-monitoring and clinical practice
Popis výsledku anglicky
Conventional glycometer systems for plasma/blood glucose monitoring are based colorimetry or static electrochemistry using a fixed input signal. The recent glucometer Linus, Wellion, Agamatrix, USA, based on wavesense dynamic electrochemistry, uses a time-varying input signal to give a more accurate glucose reading. The purpose of this study was to compare the plasma glucose (PG) readings obtained by nursing staff from glucometer Linus and PG values estimated on an approved analyzer Daytona TM, Randox,Global Medical Instrumentation, Inc., MN, USA. Glucometer Linus readings correspond to the ISO and FDA standards and meet the criteria of clinical utility. So, Linus appears to be an accurate and reliable device for PG-self-monitoring and clinical practice.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FB - Endokrinologie, diabetologie, metabolismus, výživa
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/NR7825" target="_blank" >NR7825: Stanovení glykemického indexu potravin u zdravých osob a u diabetiků 1. a 2. typu.</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Acta Diabetologica
ISSN
0940-5429
e-ISSN
—
Svazek periodika
48
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
DE - Spolková republika Německo
Počet stran výsledku
6
Strana od-do
35-40
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—