Preparative isotachophoresis with surface enhanced Raman scattering as a promising tool for clinical samples analysis
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61989592%3A15310%2F11%3A33117991" target="_blank" >RIV/61989592:15310/11:33117991 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.chroma.2010.11.025" target="_blank" >http://dx.doi.org/10.1016/j.chroma.2010.11.025</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.chroma.2010.11.025" target="_blank" >10.1016/j.chroma.2010.11.025</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Preparative isotachophoresis with surface enhanced Raman scattering as a promising tool for clinical samples analysis
Popis výsledku v původním jazyce
A surface enhanced Raman scattering (SERS) spectrometry is an interesting alternative for a rapid molecular recognition of analytes at very low concentration levels. The hyphenation of this technique with advanced separation methods enhances its potential as a detection technique. Until now, it has been hyphenated mainly with common chromatographic and electrophoretic techniques. This work demonstrates for a first time a power of preparative isotachophoresis-surface enhanced Raman scattering spectrometry (pITP-SERS) combination on the analysis of model analyte (buserelin) in a complex biological sample (urine). An off-line identification of target analyte was performed using a comparison of Raman spectra of buserelin standard with spectra obtained by the analyses of the fractions from preparative isotachophoretic runs. SERS determination of buserelin was based on the method of standard addition to minimize the matrix effects. The linearity of developed method was obtained in the concen
Název v anglickém jazyce
Preparative isotachophoresis with surface enhanced Raman scattering as a promising tool for clinical samples analysis
Popis výsledku anglicky
A surface enhanced Raman scattering (SERS) spectrometry is an interesting alternative for a rapid molecular recognition of analytes at very low concentration levels. The hyphenation of this technique with advanced separation methods enhances its potential as a detection technique. Until now, it has been hyphenated mainly with common chromatographic and electrophoretic techniques. This work demonstrates for a first time a power of preparative isotachophoresis-surface enhanced Raman scattering spectrometry (pITP-SERS) combination on the analysis of model analyte (buserelin) in a complex biological sample (urine). An off-line identification of target analyte was performed using a comparison of Raman spectra of buserelin standard with spectra obtained by the analyses of the fractions from preparative isotachophoretic runs. SERS determination of buserelin was based on the method of standard addition to minimize the matrix effects. The linearity of developed method was obtained in the concen
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Chromatography A
ISSN
0021-9673
e-ISSN
—
Svazek periodika
1218
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
6
Strana od-do
205-210
Kód UT WoS článku
000286700600003
EID výsledku v databázi Scopus
—