High sensitivity deep-UV LED-based z-cell photometric detector for capillary liquid chromatography
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F62156489%3A43210%2F18%3A43914286" target="_blank" >RIV/62156489:43210/18:43914286 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216305:26620/18:PU131551
Výsledek na webu
<a href="https://doi.org/10.1016/j.aca.2018.06.005" target="_blank" >https://doi.org/10.1016/j.aca.2018.06.005</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.aca.2018.06.005" target="_blank" >10.1016/j.aca.2018.06.005</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
High sensitivity deep-UV LED-based z-cell photometric detector for capillary liquid chromatography
Popis výsledku v původním jazyce
A new high sensitivity deep-UV LED photometric detector with a z-type flow cell (45 nL or 180 nL) for miniaturised and portable capillary liquid chromatography (LC) was designed and fabricated to overcome sensitivity limitations due to short pathlength in on-capillary detectors. The new detector has a 10 mm geometric pathlength and uses high intensity light-emitting diodes (LED) as light sources in the deep-UV range (254 nm and 280 nm). No optical reference was necessary due to the low drift in the signal. Stray light was minimized by the use of an adjustable slit with a 0.5mm pinhole. The direct relationship between absorbance and concentration was obtained using dichromate to evaluate the sensitivity and the linearity range of the detector. Performance of the miniaturised version was compared with that obtained from a commercial benchtop detector for capillary LC under the same conditions using the same optical z-cell. The miniaturised version exhibited a superior performance across all parameters, including 3 times higher effective pathlength, 4 times higher upper limit of detector linearity, and 2-5 times lower stray light levels. An application of the new detector was shown with the detection of L-dopa, L-tyrosine, norfenefrine, phenylephrine and tyramine, separated using capillary LC. The baseline noise level recorded was as low as 3.9 mu AU. Further, the detector was applied in a miniaturised capillary LC for the indirect detection of common inorganic anions. In comparison to an on-capillary LED detector applied under similar chromatographic conditions, there was a 50 times higher signal to noise (S/N) ratio.
Název v anglickém jazyce
High sensitivity deep-UV LED-based z-cell photometric detector for capillary liquid chromatography
Popis výsledku anglicky
A new high sensitivity deep-UV LED photometric detector with a z-type flow cell (45 nL or 180 nL) for miniaturised and portable capillary liquid chromatography (LC) was designed and fabricated to overcome sensitivity limitations due to short pathlength in on-capillary detectors. The new detector has a 10 mm geometric pathlength and uses high intensity light-emitting diodes (LED) as light sources in the deep-UV range (254 nm and 280 nm). No optical reference was necessary due to the low drift in the signal. Stray light was minimized by the use of an adjustable slit with a 0.5mm pinhole. The direct relationship between absorbance and concentration was obtained using dichromate to evaluate the sensitivity and the linearity range of the detector. Performance of the miniaturised version was compared with that obtained from a commercial benchtop detector for capillary LC under the same conditions using the same optical z-cell. The miniaturised version exhibited a superior performance across all parameters, including 3 times higher effective pathlength, 4 times higher upper limit of detector linearity, and 2-5 times lower stray light levels. An application of the new detector was shown with the detection of L-dopa, L-tyrosine, norfenefrine, phenylephrine and tyramine, separated using capillary LC. The baseline noise level recorded was as low as 3.9 mu AU. Further, the detector was applied in a miniaturised capillary LC for the indirect detection of common inorganic anions. In comparison to an on-capillary LED detector applied under similar chromatographic conditions, there was a 50 times higher signal to noise (S/N) ratio.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10406 - Analytical chemistry
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2018
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Analytica Chimica Acta
ISSN
0003-2670
e-ISSN
—
Svazek periodika
1032
Číslo periodika v rámci svazku
22 November
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
6
Strana od-do
197-202