Biosensor enhancement using grooved micromixers: Part II, experimental studies
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F67985882%3A_____%2F15%3A00449184" target="_blank" >RIV/67985882:_____/15:00449184 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1021/ac504360d" target="_blank" >http://dx.doi.org/10.1021/ac504360d</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1021/ac504360d" target="_blank" >10.1021/ac504360d</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Biosensor enhancement using grooved micromixers: Part II, experimental studies
Popis výsledku v původním jazyce
In this study we examine the experimental use of the staggered herringbone mixer (SHM) for the signal enhancement of a microfluidic surface plasmon resonance imaging (SPRi) affinity-based biosensor. We define the signal enhancement (Emix) as the ratio ofthe time-dependent slope of the sensor response of a SHM-based microfluidic channel and that of an unmixed channel; Emix is directly proportional to changes in the sensor sensitivity and inversely proportional to changes in the sensor limit of detection(LOD). Measurements were carried out for three SHM designs under a wide range of volumetric flow rates for two analytes: high diffusivity ssDNA and low diffusivity Escherichia coli bacteria. The experimental data collected in this study was found to exhibit a good match to that predicted by the numerical methods discussed in part I of this study. We found that Emix is dependent on the SHM groove geometry, the Péclet number Pe, and the overall microchannel length L; these dependencies ar
Název v anglickém jazyce
Biosensor enhancement using grooved micromixers: Part II, experimental studies
Popis výsledku anglicky
In this study we examine the experimental use of the staggered herringbone mixer (SHM) for the signal enhancement of a microfluidic surface plasmon resonance imaging (SPRi) affinity-based biosensor. We define the signal enhancement (Emix) as the ratio ofthe time-dependent slope of the sensor response of a SHM-based microfluidic channel and that of an unmixed channel; Emix is directly proportional to changes in the sensor sensitivity and inversely proportional to changes in the sensor limit of detection(LOD). Measurements were carried out for three SHM designs under a wide range of volumetric flow rates for two analytes: high diffusivity ssDNA and low diffusivity Escherichia coli bacteria. The experimental data collected in this study was found to exhibit a good match to that predicted by the numerical methods discussed in part I of this study. We found that Emix is dependent on the SHM groove geometry, the Péclet number Pe, and the overall microchannel length L; these dependencies ar
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GBP205%2F12%2FG118" target="_blank" >GBP205/12/G118: Nanobiofotonika pro medicínu budoucnosti</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Analytical Chemistry
ISSN
0003-2700
e-ISSN
—
Svazek periodika
87
Číslo periodika v rámci svazku
11
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
7
Strana od-do
5524-5530
Kód UT WoS článku
000355779500013
EID výsledku v databázi Scopus
2-s2.0-84930669341