Biosensor enhancement using grooved micromixers: Part I, numerical studies
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F67985882%3A_____%2F15%3A00449180" target="_blank" >RIV/67985882:_____/15:00449180 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1021/ac504359m" target="_blank" >http://dx.doi.org/10.1021/ac504359m</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1021/ac504359m" target="_blank" >10.1021/ac504359m</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Biosensor enhancement using grooved micromixers: Part I, numerical studies
Popis výsledku v původním jazyce
In this study we examine the use of the staggered herringbone mixer (SHM) to increase the efficiency of analyte delivery to a planar biosensor surface. Although there has been an extensive amount of research regarding the optimization of the SHM for mixing purposes, there has been very little work regarding the use of said micromixers for sensing purposes. Here, we use numerical methods to examine the effect of the SHM geometry on the efficiency of analyte delivery to a biosensor surface. We show the level of sensing enhancement of an SHM-based sensing chamber over that of an unmixed chamber has a strong dependence on the SHM geometry, the Péclet number, and the overall sensor length. The results presented herein are applicable to a very wide range ofbiosensor transduction mechanisms and target analytes
Název v anglickém jazyce
Biosensor enhancement using grooved micromixers: Part I, numerical studies
Popis výsledku anglicky
In this study we examine the use of the staggered herringbone mixer (SHM) to increase the efficiency of analyte delivery to a planar biosensor surface. Although there has been an extensive amount of research regarding the optimization of the SHM for mixing purposes, there has been very little work regarding the use of said micromixers for sensing purposes. Here, we use numerical methods to examine the effect of the SHM geometry on the efficiency of analyte delivery to a biosensor surface. We show the level of sensing enhancement of an SHM-based sensing chamber over that of an unmixed chamber has a strong dependence on the SHM geometry, the Péclet number, and the overall sensor length. The results presented herein are applicable to a very wide range ofbiosensor transduction mechanisms and target analytes
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GBP205%2F12%2FG118" target="_blank" >GBP205/12/G118: Nanobiofotonika pro medicínu budoucnosti</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Analytical Chemistry
ISSN
0003-2700
e-ISSN
—
Svazek periodika
87
Číslo periodika v rámci svazku
11
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
8
Strana od-do
5516-5523
Kód UT WoS článku
000355779500012
EID výsledku v databázi Scopus
2-s2.0-84930661078