Performance of label-free optical biosensors: What is figure of merit (not) telling us?
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F67985882%3A_____%2F22%3A00560103" target="_blank" >RIV/67985882:_____/22:00560103 - isvavai.cz</a>
Výsledek na webu
<a href="https://doi.org/10.1016/j.bios.2022.114426" target="_blank" >https://doi.org/10.1016/j.bios.2022.114426</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.bios.2022.114426" target="_blank" >10.1016/j.bios.2022.114426</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Performance of label-free optical biosensors: What is figure of merit (not) telling us?
Popis výsledku v původním jazyce
Here we study the analytical performance of label-free optical biosensors with respect to analyte-induced refractive index changes that can be measured by a biosensor (refractive index resolution). We present an analytical model that interrelates the refractive index resolution and the parameters of the optical platform of a biosensor. We demonstrate that the figure of merit (FOM), which has been widely used to design optical platforms of label-free optical biosensors, is not an appropriate metric to guide the design or predict the performance of label-free optical biosensors. Therefore, we propose an extended definition of FOM that addresses its limitations. We confirm the validity of the proposed approach by both numerical simulations and experiments. Finally, we show that the analytical model of the refractive index resolution not only makes it possible to predict the performance of a biosensor but also provides strategies for achieving optimal performance.
Název v anglickém jazyce
Performance of label-free optical biosensors: What is figure of merit (not) telling us?
Popis výsledku anglicky
Here we study the analytical performance of label-free optical biosensors with respect to analyte-induced refractive index changes that can be measured by a biosensor (refractive index resolution). We present an analytical model that interrelates the refractive index resolution and the parameters of the optical platform of a biosensor. We demonstrate that the figure of merit (FOM), which has been widely used to design optical platforms of label-free optical biosensors, is not an appropriate metric to guide the design or predict the performance of label-free optical biosensors. Therefore, we propose an extended definition of FOM that addresses its limitations. We confirm the validity of the proposed approach by both numerical simulations and experiments. Finally, we show that the analytical model of the refractive index resolution not only makes it possible to predict the performance of a biosensor but also provides strategies for achieving optimal performance.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10306 - Optics (including laser optics and quantum optics)
Návaznosti výsledku
Projekt
<a href="/cs/project/GX20-23787X" target="_blank" >GX20-23787X: Nové biofotonické nástroje pro studium buněčných procesů</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2022
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Biosensors and Bioelectronics
ISSN
0956-5663
e-ISSN
1873-4235
Svazek periodika
212
Číslo periodika v rámci svazku
SEP 15
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
5
Strana od-do
114426
Kód UT WoS článku
000833517400003
EID výsledku v databázi Scopus
2-s2.0-85131426220