Oxygen determination in the Ti certified reference material ERM-EB090b by instrumental photon activation analysis
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F67985891%3A_____%2F24%3A00579020" target="_blank" >RIV/67985891:_____/24:00579020 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/61389005:_____/24:00581042
Výsledek na webu
<a href="https://doi.org/10.1007/s10967-023-09260-1" target="_blank" >https://doi.org/10.1007/s10967-023-09260-1</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1007/s10967-023-09260-1" target="_blank" >10.1007/s10967-023-09260-1</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Oxygen determination in the Ti certified reference material ERM-EB090b by instrumental photon activation analysis
Popis výsledku v původním jazyce
For mechanical properties of titanium, oxygen content is important. In re-certifying O mass fraction in the Ti certified reference material ERM-EB090b, instrumental photon activation analysis (IPAA) was used as an alternative to the inert gas fusion technique. IPAA utilizes 15O measurable only via the nonspecific annihilation gamma line 511 keV. Interferences from other radionuclides can be suppressed by optimizing photon energy and irradiation-decay-counting times, and for most interfering radionuclides corrected via their specific gamma lines. The present re-evaluated IPAA results for O content 3.56 +/- 0.59 g/kg matches closely the assigned certified value 3.57 +/- 0.19 g/kg.
Název v anglickém jazyce
Oxygen determination in the Ti certified reference material ERM-EB090b by instrumental photon activation analysis
Popis výsledku anglicky
For mechanical properties of titanium, oxygen content is important. In re-certifying O mass fraction in the Ti certified reference material ERM-EB090b, instrumental photon activation analysis (IPAA) was used as an alternative to the inert gas fusion technique. IPAA utilizes 15O measurable only via the nonspecific annihilation gamma line 511 keV. Interferences from other radionuclides can be suppressed by optimizing photon energy and irradiation-decay-counting times, and for most interfering radionuclides corrected via their specific gamma lines. The present re-evaluated IPAA results for O content 3.56 +/- 0.59 g/kg matches closely the assigned certified value 3.57 +/- 0.19 g/kg.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10304 - Nuclear physics
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2024
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Radioanalytical and Nuclear Chemistry
ISSN
0236-5731
e-ISSN
1588-2780
Svazek periodika
333
Číslo periodika v rámci svazku
7
Stát vydavatele periodika
DE - Spolková republika Německo
Počet stran výsledku
6
Strana od-do
3481-3486
Kód UT WoS článku
001110395400002
EID výsledku v databázi Scopus
2-s2.0-85177794124