Validation and inter-laboratory study of selective hydride generation for fast screening of inorganic arsenic in seafood
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081715%3A_____%2F19%3A00496926" target="_blank" >RIV/68081715:_____/19:00496926 - isvavai.cz</a>
Výsledek na webu
<a href="http://hdl.handle.net/11104/0289533" target="_blank" >http://hdl.handle.net/11104/0289533</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.aca.2018.11.036" target="_blank" >10.1016/j.aca.2018.11.036</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Validation and inter-laboratory study of selective hydride generation for fast screening of inorganic arsenic in seafood
Popis výsledku v původním jazyce
This work describes an update and validation of a method of selective hydride generation (HG) with inductively coupled plasma mass spectrometry for high-throughput screening of toxic inorganic arsenic (iAs) content in seafood samples after microwave-assisted extraction with diluted nitric acid and hydrogen peroxide. High concentration of HCl (8 M) for HG along with hydrogen peroxide in samples leads to a selective conversion of iAs to volatile arsine that is released and transported to the detector, thus allowing for non-chromatographic separation of iAs from the organically bound As species that are considered less toxic or even non-toxic.
Název v anglickém jazyce
Validation and inter-laboratory study of selective hydride generation for fast screening of inorganic arsenic in seafood
Popis výsledku anglicky
This work describes an update and validation of a method of selective hydride generation (HG) with inductively coupled plasma mass spectrometry for high-throughput screening of toxic inorganic arsenic (iAs) content in seafood samples after microwave-assisted extraction with diluted nitric acid and hydrogen peroxide. High concentration of HCl (8 M) for HG along with hydrogen peroxide in samples leads to a selective conversion of iAs to volatile arsine that is released and transported to the detector, thus allowing for non-chromatographic separation of iAs from the organically bound As species that are considered less toxic or even non-toxic.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10406 - Analytical chemistry
Návaznosti výsledku
Projekt
<a href="/cs/project/LH15174" target="_blank" >LH15174: Pokročilá speciační analýza arsenu v obtížných biologických matricích</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Analytica Chimica Acta
ISSN
0003-2670
e-ISSN
—
Svazek periodika
1049
Číslo periodika v rámci svazku
FEB
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
9
Strana od-do
20-28
Kód UT WoS článku
000454789100003
EID výsledku v databázi Scopus
2-s2.0-85057229279