Helium cryopump for electron optical devices design and preliminary cryogenic testing.
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F00%3A12000043" target="_blank" >RIV/68081731:_____/00:12000043 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Helium cryopump for electron optical devices design and preliminary cryogenic testing.
Popis výsledku v původním jazyce
A helium cryopump for electron optical devices was designed and preliminary cryogenic parameters of the first pump sample were tested. Low LHe consumption should enable economical operation of the pump without recycling the evaporated He. The pumping cryopanel is cooled by normal LHe down to 4.2 K. A LN2 bath is utilized for the LHe bath thermal shielding. Two additional heat radiation shields cooled by the vapours of the consumed cryoliquids are installed to decrease the LHe and LN2 losses. Three chevron screens placed between the input office and the cryopanel are used to protect the internal parts against direct heat radiation from the parts at ambient temperature. The chevron systems were replaced by low emissivity screens in the preliniary tests.Predicted low evaporation rates of LHe and LN2 were achieved.
Název v anglickém jazyce
Helium cryopump for electron optical devices design and preliminary cryogenic testing.
Popis výsledku anglicky
A helium cryopump for electron optical devices was designed and preliminary cryogenic parameters of the first pump sample were tested. Low LHe consumption should enable economical operation of the pump without recycling the evaporated He. The pumping cryopanel is cooled by normal LHe down to 4.2 K. A LN2 bath is utilized for the LHe bath thermal shielding. Two additional heat radiation shields cooled by the vapours of the consumed cryoliquids are installed to decrease the LHe and LN2 losses. Three chevron screens placed between the input office and the cryopanel are used to protect the internal parts against direct heat radiation from the parts at ambient temperature. The chevron systems were replaced by low emissivity screens in the preliniary tests.Predicted low evaporation rates of LHe and LN2 were achieved.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BM - Fyzika pevných látek a magnetismus
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA101%2F00%2F0028" target="_blank" >GA101/00/0028: Heliová kryogenní vývěva pro elektronově optické přístroje</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2000
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of the 6th Cryogenics 2000 IIR International Conference - Refrigeration Science and Technology.
ISBN
—
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
—
Název nakladatele
Czech Committee for co-operation with IIR
Místo vydání
Praha
Místo konání akce
—
Datum konání akce
—
Typ akce podle státní příslušnosti
—
Kód UT WoS článku
—