Economical helium bath cryopump: design and testing
Popis výsledku v původním jazyce
A small UHV helium bath cryopump with low cryoliquid consumption has been designed, manufactured and tested. The cryopump is designed to keep UHV in electron optical devices without generation of disturbing electromagnetic fields or vibrations. The outervolume of its shell is 15l, the filling volumes of liquid helium (LHe) and liquid nitrogen (LN2) are 3.0 l and 3.4 l, respectively. Operating times between cryoliquid fillings are 35 days for LHe and 6 days for LN2. A pump speed of 25l/s was measured for the He gas. A lowest pressure of 1.5e-7 Pa was achieved in a test chamber and the pumping tests indicated that an even lower pressure would be achieved in a well-degassed chamber. Good agreement between the calculated and measured values both of the pump speed and the cryoliquid evaporation has been found. The system of chevron-type baffles was designed and checked by Monte Carlo simulations of molecular flow and radiative heat transfer
Název v anglickém jazyce
Economical helium bath cryopump: design and testing
Popis výsledku anglicky
A small UHV helium bath cryopump with low cryoliquid consumption has been designed, manufactured and tested. The cryopump is designed to keep UHV in electron optical devices without generation of disturbing electromagnetic fields or vibrations. The outervolume of its shell is 15l, the filling volumes of liquid helium (LHe) and liquid nitrogen (LN2) are 3.0 l and 3.4 l, respectively. Operating times between cryoliquid fillings are 35 days for LHe and 6 days for LN2. A pump speed of 25l/s was measured for the He gas. A lowest pressure of 1.5e-7 Pa was achieved in a test chamber and the pumping tests indicated that an even lower pressure would be achieved in a well-degassed chamber. Good agreement between the calculated and measured values both of the pump speed and the cryoliquid evaporation has been found. The system of chevron-type baffles was designed and checked by Monte Carlo simulations of molecular flow and radiative heat transfer
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA101%2F00%2F0028" target="_blank" >GA101/00/0028: Heliová kryogenní vývěva pro elektronově optické přístroje</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2004
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Vacuum
ISSN
0042-207X
e-ISSN
—
Svazek periodika
74
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
7
Strana od-do
77-83
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—