Absolutní frekvenční posuvy jodem stabilizovaných laserových etalonů zpusobené čistotou jodu absorpčních kyvet
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F08%3A00315092" target="_blank" >RIV/68081731:_____/08:00315092 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Absolute Frequency Shifts of Iodine-Stabilized Laser Etalons in Relation to Iodine Purity in Absorption Cells
Popis výsledku v původním jazyce
Results of measurement of purity of a set of iodine cells for laser stabilization made at our institute are presented. The purity was tested by improved method based on measurement of induced fluorescence and evaluation by the Stern-Volmer formula. Thismethod was improved by innovation of the fluorescence detection system by introducing compensation for the pumping laser spectral and power instabilities. Further the absolute frequencies of selected iodine hyperfine transitions were measured in direct laser frequency comparison performed with a set of iodine-stabilized Nd:YAG laser etalons with the reproducibility well below the kHz level. The results indicating the iodine cell purity are presented with relation to the absolute frequency shifts. This highlights the influence of iodine cell quality onto the stability and absolute frequency of lasers etalons and also shows the way towards improvements of the iodine cell manufacturing technology.
Název v anglickém jazyce
Absolute Frequency Shifts of Iodine-Stabilized Laser Etalons in Relation to Iodine Purity in Absorption Cells
Popis výsledku anglicky
Results of measurement of purity of a set of iodine cells for laser stabilization made at our institute are presented. The purity was tested by improved method based on measurement of induced fluorescence and evaluation by the Stern-Volmer formula. Thismethod was improved by innovation of the fluorescence detection system by introducing compensation for the pumping laser spectral and power instabilities. Further the absolute frequencies of selected iodine hyperfine transitions were measured in direct laser frequency comparison performed with a set of iodine-stabilized Nd:YAG laser etalons with the reproducibility well below the kHz level. The results indicating the iodine cell purity are presented with relation to the absolute frequency shifts. This highlights the influence of iodine cell quality onto the stability and absolute frequency of lasers etalons and also shows the way towards improvements of the iodine cell manufacturing technology.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BH - Optika, masery a lasery
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2008
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of the 18th IMEKO TC 2 Symposium on Photonic in measurements
ISBN
978-80-86742-24-3
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
—
Název nakladatele
ČSSF
Místo vydání
Praha
Místo konání akce
Praha
Datum konání akce
25. 8. 2008
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—