Ultra-high-frequency ECG Measurement
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F13%3A00421203" target="_blank" >RIV/68081731:_____/13:00421203 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Ultra-high-frequency ECG Measurement
Popis výsledku v původním jazyce
The aim of the present study is to introduce ultrahigh-frequency and high-dynamic-range 12-lead ECG measurements using an electromagnetically shielded environment (Faraday cage) and a battery-powered ECG monitor with sampling of 25 kHz and 24-bit resolution. To demonstrate its diagnostic contribution we present the results of 14 subjects ? 7 healthy volunteers and 7 ischemic patients during short 15-minute resting supine recordings. Both groups of subjects have the same width of the regular QRS complexwith no ST-segment abnormalities. Wide-band power envelope analysis up to 1000 Hz was applied on each QRS complex region. Artifact-free QRS complexes from approximately 500 beats were averaged with the R-wave maximum from lead V2 as a trigger. The powerenvelopes consist of narrow and compact shapes in all leads in healthy young volunteers. In ischemic patients, there is a significant expansion and splitting of frequency components and differentiation in measured leads. In this study, we
Název v anglickém jazyce
Ultra-high-frequency ECG Measurement
Popis výsledku anglicky
The aim of the present study is to introduce ultrahigh-frequency and high-dynamic-range 12-lead ECG measurements using an electromagnetically shielded environment (Faraday cage) and a battery-powered ECG monitor with sampling of 25 kHz and 24-bit resolution. To demonstrate its diagnostic contribution we present the results of 14 subjects ? 7 healthy volunteers and 7 ischemic patients during short 15-minute resting supine recordings. Both groups of subjects have the same width of the regular QRS complexwith no ST-segment abnormalities. Wide-band power envelope analysis up to 1000 Hz was applied on each QRS complex region. Artifact-free QRS complexes from approximately 500 beats were averaged with the R-wave maximum from lead V2 as a trigger. The powerenvelopes consist of narrow and compact shapes in all leads in healthy young volunteers. In ischemic patients, there is a significant expansion and splitting of frequency components and differentiation in measured leads. In this study, we
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/EE2.4.31.0016" target="_blank" >EE2.4.31.0016: MEDTECH - vzdělávací a výzkumná partnerská síť v medicíně, biomedicíně a přístrojové technice</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Computing in Cardiology /40./
ISBN
—
ISSN
2325-8861
e-ISSN
—
Počet stran výsledku
4
Strana od-do
783-786
Název nakladatele
Computing in Cardiology
Místo vydání
Zaragoza
Místo konání akce
Zaragoza
Datum konání akce
22. 9. 2013
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—