Effect of thermal radiation on low temperature measurement for various types of temperature sensor installations
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F17%3A00478280" target="_blank" >RIV/68081731:_____/17:00478280 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.18462/iir.cryo.2017.0105" target="_blank" >http://dx.doi.org/10.18462/iir.cryo.2017.0105</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.18462/iir.cryo.2017.0105" target="_blank" >10.18462/iir.cryo.2017.0105</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Effect of thermal radiation on low temperature measurement for various types of temperature sensor installations
Popis výsledku v původním jazyce
We investigated an effect of thermal radiation (300 K) on precision of low temperature measurement of a sample holder (or sample) in an UHV SEM/SPM microscope working at range down to 20 K for various ways of temperature sensor mounting. We designed a special copper casing for installation of Lake Shorentemperature sensors (CernoxTM and Si diodes) encapsulated in SD packages. The sensor body is attached direct to the casing using indium solder. The current leads pertaining to the sensor are thermally anchored but electrically insulated from the casing. Quality of sensors installation in the casing was tested for various ways of mounting on a cooled plate. Insufficient thermal anchoring of electrical wires and thermal shielding were imitated. Comparative measurement was also performed with sensors in SD package mounted withoutncasing. The experiments proved that our special design of temperature sensors casing is adequate for intended application of temperature measurement of sample holder in UHV SEM/SPM microscope. In contrary, measurement precision of unprotected sensors was absolutely unsuitable.
Název v anglickém jazyce
Effect of thermal radiation on low temperature measurement for various types of temperature sensor installations
Popis výsledku anglicky
We investigated an effect of thermal radiation (300 K) on precision of low temperature measurement of a sample holder (or sample) in an UHV SEM/SPM microscope working at range down to 20 K for various ways of temperature sensor mounting. We designed a special copper casing for installation of Lake Shorentemperature sensors (CernoxTM and Si diodes) encapsulated in SD packages. The sensor body is attached direct to the casing using indium solder. The current leads pertaining to the sensor are thermally anchored but electrically insulated from the casing. Quality of sensors installation in the casing was tested for various ways of mounting on a cooled plate. Insufficient thermal anchoring of electrical wires and thermal shielding were imitated. Comparative measurement was also performed with sensors in SD package mounted withoutncasing. The experiments proved that our special design of temperature sensors casing is adequate for intended application of temperature measurement of sample holder in UHV SEM/SPM microscope. In contrary, measurement precision of unprotected sensors was absolutely unsuitable.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20303 - Thermodynamics
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
14th Cryogenics 2017. IIR International Conference. Proceedings
ISBN
978-2-36215-022-7
ISSN
0151-1637
e-ISSN
—
Počet stran výsledku
4
Strana od-do
498-501
Název nakladatele
ICCEX
Místo vydání
Praha
Místo konání akce
Dresden
Datum konání akce
15. 5. 2017
Typ akce podle státní příslušnosti
WRD - Celosvětová akce