Label-free CARS microscopy through a multimode fiber endoscope
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68081731%3A_____%2F19%3A00510088" target="_blank" >RIV/68081731:_____/19:00510088 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.osapublishing.org/oe/fulltext.cfm?uri=oe-27-21-30055&id=421966" target="_blank" >https://www.osapublishing.org/oe/fulltext.cfm?uri=oe-27-21-30055&id=421966</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1364/OE.27.030055" target="_blank" >10.1364/OE.27.030055</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Label-free CARS microscopy through a multimode fiber endoscope
Popis výsledku v původním jazyce
Multimode fibres have recently been employed as high-resolution ultra-thin endoscopes, capable of imaging biological structures deep inside tissue in vivo. Here, we extend this technique to label-free non-linear microscopy with chemical contrast using coherent anti-Stokes Raman scattering (CARS) through a multimode fibre endoscope, which opens up new avenues for instant and in-situ diagnosis of potentially malignant tissue. We use a commercial 125 mu m diameter, 0.29 NA GRIN fibre, and wavefront shaping on an SLM is used to create foci that are scanned behind the fibre facet across the sample. The chemical selectivity is demonstrated by imaging 2 mu m polystyrene and 2.5 mu m PMMA beads with per pixel integration time as low as 1 ms for epi-detection.
Název v anglickém jazyce
Label-free CARS microscopy through a multimode fiber endoscope
Popis výsledku anglicky
Multimode fibres have recently been employed as high-resolution ultra-thin endoscopes, capable of imaging biological structures deep inside tissue in vivo. Here, we extend this technique to label-free non-linear microscopy with chemical contrast using coherent anti-Stokes Raman scattering (CARS) through a multimode fibre endoscope, which opens up new avenues for instant and in-situ diagnosis of potentially malignant tissue. We use a commercial 125 mu m diameter, 0.29 NA GRIN fibre, and wavefront shaping on an SLM is used to create foci that are scanned behind the fibre facet across the sample. The chemical selectivity is demonstrated by imaging 2 mu m polystyrene and 2.5 mu m PMMA beads with per pixel integration time as low as 1 ms for epi-detection.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10306 - Optics (including laser optics and quantum optics)
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Optics Express
ISSN
1094-4087
e-ISSN
—
Svazek periodika
27
Číslo periodika v rámci svazku
21
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
12
Strana od-do
30055-30066
Kód UT WoS článku
000489954500053
EID výsledku v databázi Scopus
2-s2.0-85073542251