Comparison of Lu3Al5O12:Pr3+ and Bi4Ge3O12 scintillators for gamma-ray detection
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68378271%3A_____%2F12%3A00382206" target="_blank" >RIV/68378271:_____/12:00382206 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.radmeas.2011.08.023" target="_blank" >http://dx.doi.org/10.1016/j.radmeas.2011.08.023</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.radmeas.2011.08.023" target="_blank" >10.1016/j.radmeas.2011.08.023</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Comparison of Lu3Al5O12:Pr3+ and Bi4Ge3O12 scintillators for gamma-ray detection
Popis výsledku v původním jazyce
The scintillation properties of Lu3Al5O12:Pr3? (LuAG:Pr) single crystal grown by the Czochralski method with praseodymium concentration of 0.19 mol% were investigated. For a comparison, a good quality Bi4Ge3O12 (BGO) single crystal grown by Bridgman method was also studied. The light yield and energy resolution were measured using photomultiplier tube (XP5200B PMT) readout. Moderate light yield of 15,900 photons per MeV was measured for the LuAG:Pr(0.19%) crystal. For 662 keV gamma rays (137Cs source),an energy resolution of 6.5% obtained for LuAG:Pr(0.19%) is much better than that of 9.0% obtained for BGO. The light yield non-proportionality and energy resolution versus energy of gamma rays were measured and the intrinsic resolution of the crystals was determined after correcting the measured energy resolution for PMT statistics.
Název v anglickém jazyce
Comparison of Lu3Al5O12:Pr3+ and Bi4Ge3O12 scintillators for gamma-ray detection
Popis výsledku anglicky
The scintillation properties of Lu3Al5O12:Pr3? (LuAG:Pr) single crystal grown by the Czochralski method with praseodymium concentration of 0.19 mol% were investigated. For a comparison, a good quality Bi4Ge3O12 (BGO) single crystal grown by Bridgman method was also studied. The light yield and energy resolution were measured using photomultiplier tube (XP5200B PMT) readout. Moderate light yield of 15,900 photons per MeV was measured for the LuAG:Pr(0.19%) crystal. For 662 keV gamma rays (137Cs source),an energy resolution of 6.5% obtained for LuAG:Pr(0.19%) is much better than that of 9.0% obtained for BGO. The light yield non-proportionality and energy resolution versus energy of gamma rays were measured and the intrinsic resolution of the crystals was determined after correcting the measured energy resolution for PMT statistics.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BM - Fyzika pevných látek a magnetismus
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA202%2F08%2F0893" target="_blank" >GA202/08/0893: Nové materiály a technologie pro přípravu tenkovrstvých scintilátorů</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Measurements
ISSN
1350-4487
e-ISSN
—
Svazek periodika
47
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
6
Strana od-do
1-5
Kód UT WoS článku
00300810900001