Compact high energy x-ray spectrometer based on forward Compton scattering for high intensity laser plasma experiments
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68378271%3A_____%2F18%3A00501825" target="_blank" >RIV/68378271:_____/18:00501825 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1063/1.5040979" target="_blank" >http://dx.doi.org/10.1063/1.5040979</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1063/1.5040979" target="_blank" >10.1063/1.5040979</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Compact high energy x-ray spectrometer based on forward Compton scattering for high intensity laser plasma experiments
Popis výsledku v původním jazyce
This article describes the design and presents recent results from testing and calibration of a forward Compton scattering high energy X-ray spectrometer. The calibration was performed using a bremsstrahlung source on the photon scattering facility at the gamma Electron linac for beams with high brilliance and low emittance accelerator at Helmholtz-Zentrum Dresden-Rossendorf, which provides high energy X-ray photons with energies up to 18 MeV. The calibration was conducted at different bremsstrahlung end point energies-10.5, 13, 15, and 18 MeV. Experimental spectra show a systematic increase in the maximum energy, photon temperature, and flux. The spectrometer is effective for an energy range of 4-20 MeV with 20% - 30% energy resolution. The spectrometer operates in low vacuum with pressure less than 0.1 mbar. Experimental tests showed that operating such a spectrometer in air causes a spuriously enhanced high energy signal due to Compton scattering of photons within air.
Název v anglickém jazyce
Compact high energy x-ray spectrometer based on forward Compton scattering for high intensity laser plasma experiments
Popis výsledku anglicky
This article describes the design and presents recent results from testing and calibration of a forward Compton scattering high energy X-ray spectrometer. The calibration was performed using a bremsstrahlung source on the photon scattering facility at the gamma Electron linac for beams with high brilliance and low emittance accelerator at Helmholtz-Zentrum Dresden-Rossendorf, which provides high energy X-ray photons with energies up to 18 MeV. The calibration was conducted at different bremsstrahlung end point energies-10.5, 13, 15, and 18 MeV. Experimental spectra show a systematic increase in the maximum energy, photon temperature, and flux. The spectrometer is effective for an energy range of 4-20 MeV with 20% - 30% energy resolution. The spectrometer operates in low vacuum with pressure less than 0.1 mbar. Experimental tests showed that operating such a spectrometer in air causes a spuriously enhanced high energy signal due to Compton scattering of photons within air.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10305 - Fluids and plasma physics (including surface physics)
Návaznosti výsledku
Projekt
<a href="/cs/project/LQ1606" target="_blank" >LQ1606: ELI Beamlines: Mezinárodní centrum Excelemce</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2018
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Review of Scientific Instruments
ISSN
0034-6748
e-ISSN
—
Svazek periodika
89
Číslo periodika v rámci svazku
8
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
8
Strana od-do
—
Kód UT WoS článku
000443720400089
EID výsledku v databázi Scopus