Laboratory testing of the Leica AT401 laser tracker
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21110%2F16%3A00242810" target="_blank" >RIV/68407700:21110/16:00242810 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.14311/AP.2016.56.0088" target="_blank" >http://dx.doi.org/10.14311/AP.2016.56.0088</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.14311/AP.2016.56.0088" target="_blank" >10.14311/AP.2016.56.0088</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Laboratory testing of the Leica AT401 laser tracker
Popis výsledku v původním jazyce
This paper describes laboratory tests on a Leica AT401laser tracker. As the newer Leica AT402 model also uses the same firmware package, most of the results should also be valid for this device. First, we present the instrument’s firmware errors and the software used for testing. The ASME B89.4.19-2006 standard for testing laser trackers is briefly presented. The warm-up effect of the instrument is inspected with respect to both angle measurement and distance measurement. The absolute distance meter (ADM) is compared with a laboratory interferometer on a 30-meter long rail and also on a bench with automated movement of the carriage of the reflector. A time series of measurements for determining the additive constant is evaluated. A simple test of the stability of the distance measurement in field conditions is introduced. Most of the tests were carried out at the Research Institute of Geodesy, Topography and Cartography (RIGTC) and at the Faculty of Civil Engineering (FCE) of the Czech Technical University in Prague (CTU).
Název v anglickém jazyce
Laboratory testing of the Leica AT401 laser tracker
Popis výsledku anglicky
This paper describes laboratory tests on a Leica AT401laser tracker. As the newer Leica AT402 model also uses the same firmware package, most of the results should also be valid for this device. First, we present the instrument’s firmware errors and the software used for testing. The ASME B89.4.19-2006 standard for testing laser trackers is briefly presented. The warm-up effect of the instrument is inspected with respect to both angle measurement and distance measurement. The absolute distance meter (ADM) is compared with a laboratory interferometer on a 30-meter long rail and also on a bench with automated movement of the carriage of the reflector. A time series of measurements for determining the additive constant is evaluated. A simple test of the stability of the distance measurement in field conditions is introduced. Most of the tests were carried out at the Research Institute of Geodesy, Topography and Cartography (RIGTC) and at the Faculty of Civil Engineering (FCE) of the Czech Technical University in Prague (CTU).
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
DE - Zemský magnetismus, geodesie, geografie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Acta Polytechnica
ISSN
1805-2363
e-ISSN
—
Svazek periodika
56
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
11
Strana od-do
88-98
Kód UT WoS článku
—
EID výsledku v databázi Scopus
2-s2.0-84966474689