Vector Feedback Homogeneity and Inner Layout Influence on Fluxgate Sensor Parameters
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F15%3A00221715" target="_blank" >RIV/68407700:21230/15:00221715 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1109/TIM.2014.2362831" target="_blank" >http://dx.doi.org/10.1109/TIM.2014.2362831</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/TIM.2014.2362831" target="_blank" >10.1109/TIM.2014.2362831</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Vector Feedback Homogeneity and Inner Layout Influence on Fluxgate Sensor Parameters
Popis výsledku v původním jazyce
Vector feedback is a concept which can significantly improve linearity and stability of a magnetic field sensor. The feedback coils effectively cancel the measured magnetic field in the inner volume of the triaxial sensor. Thus, in case of fluxgates, itsuppresses one possible source of nonlinearity? cross-field sensitivity error. The triaxial sensor axes orthogonality should be primarily defined by the orientation of the feedback coils, while the sensitivities are defined by feedback coil constants. The influence of the homogeneity of the feedback field and the influence of the sensor inner layout on calibration parameters of a vectorially compensated triaxial fluxgate magnetometer are presented.
Název v anglickém jazyce
Vector Feedback Homogeneity and Inner Layout Influence on Fluxgate Sensor Parameters
Popis výsledku anglicky
Vector feedback is a concept which can significantly improve linearity and stability of a magnetic field sensor. The feedback coils effectively cancel the measured magnetic field in the inner volume of the triaxial sensor. Thus, in case of fluxgates, itsuppresses one possible source of nonlinearity? cross-field sensitivity error. The triaxial sensor axes orthogonality should be primarily defined by the orientation of the feedback coils, while the sensitivities are defined by feedback coil constants. The influence of the homogeneity of the feedback field and the influence of the sensor inner layout on calibration parameters of a vectorially compensated triaxial fluxgate magnetometer are presented.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JB - Senzory, čidla, měření a regulace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GP13-39088P" target="_blank" >GP13-39088P: Studium vlivu konstrukce senzoru na parametry vektorového magnetometru metodou rozšířené skalární kalibrace</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
IEEE Transactions on Instrumentation and Measurement
ISSN
0018-9456
e-ISSN
—
Svazek periodika
64
Číslo periodika v rámci svazku
5
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
7
Strana od-do
1293-1299
Kód UT WoS článku
000352492900023
EID výsledku v databázi Scopus
2-s2.0-84908368734