International comparison of current transformer calibration systems up to 10 kA at 50 Hz frequency
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F16%3A00306615" target="_blank" >RIV/68407700:21230/16:00306615 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1109/CPEM.2016.7540493" target="_blank" >http://dx.doi.org/10.1109/CPEM.2016.7540493</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/CPEM.2016.7540493" target="_blank" >10.1109/CPEM.2016.7540493</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
International comparison of current transformer calibration systems up to 10 kA at 50 Hz frequency
Popis výsledku v původním jazyce
Current transformers (CTs) are precision devices that scale high currents down to values that can be easily handled by measurement equipment. To support CT applications in revenue metering, a comparison on AC current transformer calibration systems was performed among 15 European national metrology institutes using a precision CT as the travelling device. The first comparison results for the transformation ratios (4, 5, 6, 8, 10) kA/5 A of the travelling CT at nominal burden of 5 VA and 15 VA indicate good agreement between the participating laboratories. The main differences are found for phase displacement, at least partly caused by the instability of the traveling standard.
Název v anglickém jazyce
International comparison of current transformer calibration systems up to 10 kA at 50 Hz frequency
Popis výsledku anglicky
Current transformers (CTs) are precision devices that scale high currents down to values that can be easily handled by measurement equipment. To support CT applications in revenue metering, a comparison on AC current transformer calibration systems was performed among 15 European national metrology institutes using a precision CT as the travelling device. The first comparison results for the transformation ratios (4, 5, 6, 8, 10) kA/5 A of the travelling CT at nominal burden of 5 VA and 15 VA indicate good agreement between the participating laboratories. The main differences are found for phase displacement, at least partly caused by the instability of the traveling standard.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Conference on Precision Electromagnetic Measurements (CPEM 2016), 2016
ISBN
978-1-4673-9134-4
ISSN
—
e-ISSN
—
Počet stran výsledku
3
Strana od-do
—
Název nakladatele
IEEE Conference Publications
Místo vydání
Piscataway
Místo konání akce
Ottawa
Datum konání akce
10. 7. 2016
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000383955100043