A Single Helical Interstitial Applicator for Microwave Hyperthermia
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21230%2F25%3A00384714" target="_blank" >RIV/68407700:21230/25:00384714 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
A Single Helical Interstitial Applicator for Microwave Hyperthermia
Popis výsledku v původním jazyce
This article presents the design and experimental evaluation of a single helical interstitial applicator for localised microwave hyperthermia treatment. The applicator operates at a frequency of 2.45 GHz and is intended to treat small tumours with a 2–3 cm diameter located under the skin. The study includes numerical simulations in Sim4Life focusing on the distribution of the specific absorption rate (SAR) and the reflection coefficient |s11|. The designed applicator was manufactured and tested using an agar phantom with dielectric and thermal properties similar to human muscle tissue. Experimental results confirm efficient energy delivery with a measured reflection coefficient of -46.5 dB and a controlled thermal response. This research supports the potential of single-antenna hyperthermia for interstitial cancer treatment. At the same time, the optimised applicator can be used in the design of a more complex hyperthermia system.
Název v anglickém jazyce
A Single Helical Interstitial Applicator for Microwave Hyperthermia
Popis výsledku anglicky
This article presents the design and experimental evaluation of a single helical interstitial applicator for localised microwave hyperthermia treatment. The applicator operates at a frequency of 2.45 GHz and is intended to treat small tumours with a 2–3 cm diameter located under the skin. The study includes numerical simulations in Sim4Life focusing on the distribution of the specific absorption rate (SAR) and the reflection coefficient |s11|. The designed applicator was manufactured and tested using an agar phantom with dielectric and thermal properties similar to human muscle tissue. Experimental results confirm efficient energy delivery with a measured reflection coefficient of -46.5 dB and a controlled thermal response. This research supports the potential of single-antenna hyperthermia for interstitial cancer treatment. At the same time, the optimised applicator can be used in the design of a more complex hyperthermia system.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2025
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of the International Student Scientific Conference POSTER – 29/2025
ISBN
978-80-01-07422-0
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
16-20
Název nakladatele
Fakulta elektrotechnická
Místo vydání
Praha
Místo konání akce
Praha
Datum konání akce
22. 5. 2025
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—