EURADOS intercomparison exercise on MC modelling for the in-vivo monitoring of AM-241 in skull phantoms (Part II and III).
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21340%2F15%3A00238824" target="_blank" >RIV/68407700:21340/15:00238824 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/86652052:_____/15:#0000369
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.radphyschem.2015.04.009" target="_blank" >http://dx.doi.org/10.1016/j.radphyschem.2015.04.009</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.radphyschem.2015.04.009" target="_blank" >10.1016/j.radphyschem.2015.04.009</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
EURADOS intercomparison exercise on MC modelling for the in-vivo monitoring of AM-241 in skull phantoms (Part II and III).
Popis výsledku v původním jazyce
An intercomparison on in-vivo monitoring for determination of Am-241 in three skull phantoms was launched by EURADOS in 2011. The project focused on measurement and estimation of the activity of Am-241 in the human skull. Three human skull phantoms of different complexity were used. A Monte Carlo (MC) intercomparison exercise with the voxel representations of the physical phantom was launched additionally in September of 2012. The main goals of the action were (1) to investigate the different methodologies for developing MC calibrations that might arise from a complex radiological assessment and (2) to compare individual approaches of the participating laboratories in order to determine international guidance for best practice. The MC exercise consisted of three tasks with increasing difficulty, in order to test the extent of skills needed by the participating laboratory. The first task was to simulate a given detector and a well-defined semi-skull phantom. The second and third tasks presented in this paper-introduced more complex simulations with individual geometry and real detector modelling. The paper provides an overview of the participant’s results, analyses of the observed issues concerning tasks two and three, and a general evaluation of the whole project.
Název v anglickém jazyce
EURADOS intercomparison exercise on MC modelling for the in-vivo monitoring of AM-241 in skull phantoms (Part II and III).
Popis výsledku anglicky
An intercomparison on in-vivo monitoring for determination of Am-241 in three skull phantoms was launched by EURADOS in 2011. The project focused on measurement and estimation of the activity of Am-241 in the human skull. Three human skull phantoms of different complexity were used. A Monte Carlo (MC) intercomparison exercise with the voxel representations of the physical phantom was launched additionally in September of 2012. The main goals of the action were (1) to investigate the different methodologies for developing MC calibrations that might arise from a complex radiological assessment and (2) to compare individual approaches of the participating laboratories in order to determine international guidance for best practice. The MC exercise consisted of three tasks with increasing difficulty, in order to test the extent of skills needed by the participating laboratory. The first task was to simulate a given detector and a well-defined semi-skull phantom. The second and third tasks presented in this paper-introduced more complex simulations with individual geometry and real detector modelling. The paper provides an overview of the participant’s results, analyses of the observed issues concerning tasks two and three, and a general evaluation of the whole project.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
DL - Jaderné odpady, radioaktivní znečištění a kontrola
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/VG20122014093" target="_blank" >VG20122014093: Systém pro měření vnitřní kontaminace po havárii JEZ zaměřený na štítné žlázy u dětí a kontaminaci transurany</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Physics and Chemistry
ISSN
0969-806X
e-ISSN
—
Svazek periodika
113
Číslo periodika v rámci svazku
AUG
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
13
Strana od-do
59-71
Kód UT WoS článku
000356196600011
EID výsledku v databázi Scopus
2-s2.0-84929340515