Table-Top Water-Window Microscope Using a Capillary Discharge Plasma Source with Spatial Resolution 75 nm
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21340%2F20%3A00346042" target="_blank" >RIV/68407700:21340/20:00346042 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/68407700:21460/20:00346042
Výsledek na webu
<a href="https://doi.org/10.3390/app10186373" target="_blank" >https://doi.org/10.3390/app10186373</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.3390/app10186373" target="_blank" >10.3390/app10186373</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Table-Top Water-Window Microscope Using a Capillary Discharge Plasma Source with Spatial Resolution 75 nm
Popis výsledku v původním jazyce
We present a design of a compact transmission water-window microscope based on the Z-pinching capillary discharge nitrogen plasma source. The microscope operates at wavelength of 2.88 nm (430 eV), and with its table-top dimensions provides an alternative to large-scale soft X-ray (SXR) microscope systems based on synchrotrons and free-electron lasers. The emitted soft X-ray radiation is filtered by a titanium foil and focused by an ellipsoidal condenser mirror into the sample plane. A Fresnel zone plate was used to create a transmission image of the sample onto a charge-coupled device (CCD) camera. To assess the resolution of the microscope, we imaged a standard sample-copper mesh. The spatial resolution of the microscope is 75 nm at half-pitch, calculated via a 10-90% intensity knife-edge test. The applicability of the microscope is demonstrated by the imaging of green algae-Desmodesmus communis. This paper describes the principle of capillary discharge source, design of the microscope, and experimental imaging results of Cu mesh and biological sample.
Název v anglickém jazyce
Table-Top Water-Window Microscope Using a Capillary Discharge Plasma Source with Spatial Resolution 75 nm
Popis výsledku anglicky
We present a design of a compact transmission water-window microscope based on the Z-pinching capillary discharge nitrogen plasma source. The microscope operates at wavelength of 2.88 nm (430 eV), and with its table-top dimensions provides an alternative to large-scale soft X-ray (SXR) microscope systems based on synchrotrons and free-electron lasers. The emitted soft X-ray radiation is filtered by a titanium foil and focused by an ellipsoidal condenser mirror into the sample plane. A Fresnel zone plate was used to create a transmission image of the sample onto a charge-coupled device (CCD) camera. To assess the resolution of the microscope, we imaged a standard sample-copper mesh. The spatial resolution of the microscope is 75 nm at half-pitch, calculated via a 10-90% intensity knife-edge test. The applicability of the microscope is demonstrated by the imaging of green algae-Desmodesmus communis. This paper describes the principle of capillary discharge source, design of the microscope, and experimental imaging results of Cu mesh and biological sample.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10606 - Microbiology
Návaznosti výsledku
Projekt
<a href="/cs/project/LTT17015" target="_blank" >LTT17015: Výzkum v rámci Mezinárodního centra hustého magnetizovaného plazmatu</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2020
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Applied Sciences
ISSN
2076-3417
e-ISSN
2076-3417
Svazek periodika
10
Číslo periodika v rámci svazku
18
Stát vydavatele periodika
CH - Švýcarská konfederace
Počet stran výsledku
8
Strana od-do
1-8
Kód UT WoS článku
000581385600001
EID výsledku v databázi Scopus
2-s2.0-85091816425