Patients connected to a high-frequency oscillatory ventilator should be paralysed or highly sedated so that the ventilator is able to work. The aim of the study is to design a special "Demand Flow System" allowing spontaneous breathing in patients ventilated using a high-frequency ventilator. The principle of the device is a fast and exact monitoring of pressure in the ventilatory circuit, followed by a separation of the pressure signal components generated by the oscillatory ventilator and by patient'sbreathing activity. The latest mentioned signal is used for a real-time control of the extra flow injected into the ventilatory circuit so that the extra flow could compensate the pressure changes generated by the patient. The system reduces imposed work of breathing by 80 % and it allows spontaneous breathing during high-frequency oscillatory ventilation.
Název v anglickém jazyce
Allowing Spontaneous Breathing During High-Frequency Oscillatory Ventilation
Popis výsledku anglicky
Patients connected to a high-frequency oscillatory ventilator should be paralysed or highly sedated so that the ventilator is able to work. The aim of the study is to design a special "Demand Flow System" allowing spontaneous breathing in patients ventilated using a high-frequency ventilator. The principle of the device is a fast and exact monitoring of pressure in the ventilatory circuit, followed by a separation of the pressure signal components generated by the oscillatory ventilator and by patient'sbreathing activity. The latest mentioned signal is used for a real-time control of the extra flow injected into the ventilatory circuit so that the extra flow could compensate the pressure changes generated by the patient. The system reduces imposed work of breathing by 80 % and it allows spontaneous breathing during high-frequency oscillatory ventilation.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2008
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Lékař a technika
ISSN
0301-5491
e-ISSN
—
Svazek periodika
38
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
3
Strana od-do
—
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—