Model-Based Method and Instrumentation for Noninvasive Identification of Local Ischemic Lesions in the Heart
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F10%3A00197362" target="_blank" >RIV/68407700:21460/10:00197362 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Model-Based Method and Instrumentation for Noninvasive Identification of Local Ischemic Lesions in the Heart
Popis výsledku v původním jazyce
An inverse method for identification of possible local ischemic lesions is presented. Body surface potentials recorded in conditions with and without manifestation of ischemia were used to evaluate changes in time integrals of surface potentials caused by changed repolarization of ischemic myocardial cells. Surface distribution of integral differences together with information on torso volume conductor was used to find an equivalent electric generator with one or two dipoles representing the ischemic lesions. Compact system for multichannel ECG measurement and evaluation with an intelligent battery powered data acquisition unit that enables simultaneous recording of up to 128 torso leads was developed. Active electrodes and optical connection to the controlling computer were used to achieve high signal quality and patient safety. Software for ECG acquisition and evaluation of isointegral surface maps was applied for location of one or two dipoles representing myocardium areas with chan
Název v anglickém jazyce
Model-Based Method and Instrumentation for Noninvasive Identification of Local Ischemic Lesions in the Heart
Popis výsledku anglicky
An inverse method for identification of possible local ischemic lesions is presented. Body surface potentials recorded in conditions with and without manifestation of ischemia were used to evaluate changes in time integrals of surface potentials caused by changed repolarization of ischemic myocardial cells. Surface distribution of integral differences together with information on torso volume conductor was used to find an equivalent electric generator with one or two dipoles representing the ischemic lesions. Compact system for multichannel ECG measurement and evaluation with an intelligent battery powered data acquisition unit that enables simultaneous recording of up to 128 torso leads was developed. Active electrodes and optical connection to the controlling computer were used to achieve high signal quality and patient safety. Software for ECG acquisition and evaluation of isointegral surface maps was applied for location of one or two dipoles representing myocardium areas with chan
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2010
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
6th World Congress of Biomechanics (WCB 2010), 1 - 6 August 2010, Singapore
ISBN
978-3-642-14514-8
ISSN
1680-0737
e-ISSN
—
Počet stran výsledku
4
Strana od-do
1403-1406
Název nakladatele
Springer
Místo vydání
New York
Místo konání akce
Singapur
Datum konání akce
1. 8. 2010
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000300975300372