Impact of the Number of ECG Leads on the Noninvasive Localization of Premature Ventricular Activity
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F17%3A00318735" target="_blank" >RIV/68407700:21460/17:00318735 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Impact of the Number of ECG Leads on the Noninvasive Localization of Premature Ventricular Activity
Popis výsledku v původním jazyce
Body surface potentials from 96 ECG leads, individual torso models based on CT scans and inverse solution to one dipole were used for noninvasive localization of the focus of ectopic ventricular activity in five patients with frequent premature complexes. The influence of the number of ECG leads used for the inverse solution on the accuracy and stability of the localization was studied using both, homogeneous or inhomogeneous torso models. When the homogeneous torso models were used, the localization of the focus was quite stable and reduction to 64 or 48 leads preserved the acceptable accuracy of the solution. When inhomogeneous torso models were used in the inverse computations, already a reduction from 96 to 64 used ECG leads resulted in several incorrect localizations.
Název v anglickém jazyce
Impact of the Number of ECG Leads on the Noninvasive Localization of Premature Ventricular Activity
Popis výsledku anglicky
Body surface potentials from 96 ECG leads, individual torso models based on CT scans and inverse solution to one dipole were used for noninvasive localization of the focus of ectopic ventricular activity in five patients with frequent premature complexes. The influence of the number of ECG leads used for the inverse solution on the accuracy and stability of the localization was studied using both, homogeneous or inhomogeneous torso models. When the homogeneous torso models were used, the localization of the focus was quite stable and reduction to 64 or 48 leads preserved the acceptable accuracy of the solution. When inhomogeneous torso models were used in the inverse computations, already a reduction from 96 to 64 used ECG leads resulted in several incorrect localizations.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
30201 - Cardiac and Cardiovascular systems
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Trendy v biomedicínském inženýrství 2017
ISBN
978-80-248-4099-4
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
1-4
Název nakladatele
VŠB-TUO
Místo vydání
Ostrava
Místo konání akce
Horní Lomná
Datum konání akce
25. 9. 2017
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—