Pressure Cost of Volume Delivery In HFOV at Different Ventilatory Frequencies
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F13%3A00210166" target="_blank" >RIV/68407700:21460/13:00210166 - isvavai.cz</a>
Výsledek na webu
<a href="http://www.degruyter.com/view/j/bmte.2013.58.issue-s1-N/bmt-2013-4365/bmt-2013-4365.xml?rskey=geRT6Y&result=2" target="_blank" >http://www.degruyter.com/view/j/bmte.2013.58.issue-s1-N/bmt-2013-4365/bmt-2013-4365.xml?rskey=geRT6Y&result=2</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1515/bmt-2013-4365" target="_blank" >10.1515/bmt-2013-4365</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Pressure Cost of Volume Delivery In HFOV at Different Ventilatory Frequencies
Popis výsledku v původním jazyce
Ventilatory pressures and tidal volume were measured during high frequency oscillatory ventilation in the ventilatory circuit. The circuit consisted of ventilator 3100 with patient circuit, lung simulator 5600i and monitor Florian for measuring of tidalvolume. A special device for synchronous measurement of pressure at three places was used for analysis of pressure waves in the ventilatory circuit. Pressure amplitudes and tidal volume was measured for different values of alveolar compliance that was set on the model. Measurement was repeated for ventilatory frequencies 3, 5, 7 and 10 Hz. Ventilatory parameters were held constant during the experiment. We have found that change of alveolar compliance has minimal effect on the pressure cost of ventilation.
Název v anglickém jazyce
Pressure Cost of Volume Delivery In HFOV at Different Ventilatory Frequencies
Popis výsledku anglicky
Ventilatory pressures and tidal volume were measured during high frequency oscillatory ventilation in the ventilatory circuit. The circuit consisted of ventilator 3100 with patient circuit, lung simulator 5600i and monitor Florian for measuring of tidalvolume. A special device for synchronous measurement of pressure at three places was used for analysis of pressure waves in the ventilatory circuit. Pressure amplitudes and tidal volume was measured for different values of alveolar compliance that was set on the model. Measurement was repeated for ventilatory frequencies 3, 5, 7 and 10 Hz. Ventilatory parameters were held constant during the experiment. We have found that change of alveolar compliance has minimal effect on the pressure cost of ventilation.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Biomedizinische Technik / Biomedical Engineering
ISSN
0013-5585
e-ISSN
—
Svazek periodika
58
Číslo periodika v rámci svazku
S1
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
2
Strana od-do
—
Kód UT WoS článku
000329463600427
EID výsledku v databázi Scopus
—