Dependency of Tidal Volume on Mean Airway Pressure in High-Frequency Oscillatory Ventilation
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F17%3A00305982" target="_blank" >RIV/68407700:21460/17:00305982 - isvavai.cz</a>
Výsledek na webu
<a href="http://users.utcluj.ro/~simona/conf/L28.pdf" target="_blank" >http://users.utcluj.ro/~simona/conf/L28.pdf</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1007/978-3-319-52875-5_39" target="_blank" >10.1007/978-3-319-52875-5_39</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Dependency of Tidal Volume on Mean Airway Pressure in High-Frequency Oscillatory Ventilation
Popis výsledku v původním jazyce
Recent studies have shown that there exists a dependence of tidal volume on a mean airway pressure in high- frequency oscillatory ventilation. Objective of the present study was to examine if the dependence is caused by changing the mean airway pressure value itself or if it is caused by a change of the mechanical properties of the respiratory system according to a change of the mean airway pressure. Ventilator 3100B and rigid respiratory system models of volumes 25 L, 35 L, 50 L and 120 L with constant mechanical properties were used in the experimental circuit. Volume flow and tidal volume was measured with custom-made orifice and 2100 Respiration Monitor. Mean airway pressure was increased from baseline value by 4 cmH2O every 1.5 min. It was shown that there is a statistically significant relationship between tidal volume and the mean airway pressure in constant mechanical properties of the respiratory system.
Název v anglickém jazyce
Dependency of Tidal Volume on Mean Airway Pressure in High-Frequency Oscillatory Ventilation
Popis výsledku anglicky
Recent studies have shown that there exists a dependence of tidal volume on a mean airway pressure in high- frequency oscillatory ventilation. Objective of the present study was to examine if the dependence is caused by changing the mean airway pressure value itself or if it is caused by a change of the mechanical properties of the respiratory system according to a change of the mean airway pressure. Ventilator 3100B and rigid respiratory system models of volumes 25 L, 35 L, 50 L and 120 L with constant mechanical properties were used in the experimental circuit. Volume flow and tidal volume was measured with custom-made orifice and 2100 Respiration Monitor. Mean airway pressure was increased from baseline value by 4 cmH2O every 1.5 min. It was shown that there is a statistically significant relationship between tidal volume and the mean airway pressure in constant mechanical properties of the respiratory system.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20601 - Medical engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
MediTech 2016 Proceedings
ISBN
978-3-319-52874-8
ISSN
1680-0737
e-ISSN
—
Počet stran výsledku
4
Strana od-do
177-180
Název nakladatele
Technical University
Místo vydání
Cluj-Napoca
Místo konání akce
Cluj-Napoca
Datum konání akce
12. 10. 2016
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000426009100039