Design Model for Risk Assessment for Home-Care Lung Ventilation
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F19%3A00321950" target="_blank" >RIV/68407700:21460/19:00321950 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1007/978-981-10-9023-3_71" target="_blank" >http://dx.doi.org/10.1007/978-981-10-9023-3_71</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1007/978-981-10-9023-3_71" target="_blank" >10.1007/978-981-10-9023-3_71</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Design Model for Risk Assessment for Home-Care Lung Ventilation
Popis výsledku v původním jazyce
Introduction: The study consists of two objectives. The first objective is to assess the risks associated with the whole process of treatment using home-care lung ventilation (HLV). The main source for describing such risks were foreign studies. The other objective is to compare risk assessment methods based on the current situation in the Czech Republic and in the world. Methods: The selected methods—Failure Mode and Effects Analysis (FMEA), Health Failure Mode and Effects Analysis (HFMEA), Fault Tree Analysis (FTA) and Root Cause Analysis (RCA)—were compared using comparative analysis, and subsequently examined, in the specific area of home-care lung ventilation. Results: The final objective is to design a model for risk assessment in the field of home mechanical ventilation based on the existing knowledge. This model includes both a prospective and a retrospective analysis. Their appropriateness of use is documented in this paper; according to their specificities they are used in a variety of areas. Discussion: Implementation of risk analyses in this area can contribute to increasing the safety and quality of provided care, and at the same time help attract more support to the home-care lung ventilation program.
Název v anglickém jazyce
Design Model for Risk Assessment for Home-Care Lung Ventilation
Popis výsledku anglicky
Introduction: The study consists of two objectives. The first objective is to assess the risks associated with the whole process of treatment using home-care lung ventilation (HLV). The main source for describing such risks were foreign studies. The other objective is to compare risk assessment methods based on the current situation in the Czech Republic and in the world. Methods: The selected methods—Failure Mode and Effects Analysis (FMEA), Health Failure Mode and Effects Analysis (HFMEA), Fault Tree Analysis (FTA) and Root Cause Analysis (RCA)—were compared using comparative analysis, and subsequently examined, in the specific area of home-care lung ventilation. Results: The final objective is to design a model for risk assessment in the field of home mechanical ventilation based on the existing knowledge. This model includes both a prospective and a retrospective analysis. Their appropriateness of use is documented in this paper; according to their specificities they are used in a variety of areas. Discussion: Implementation of risk analyses in this area can contribute to increasing the safety and quality of provided care, and at the same time help attract more support to the home-care lung ventilation program.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20601 - Medical engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
World Congress on Medical Physics and Biomedical Engineering 2018 (Vol. 3)
ISBN
978-981-10-9022-6
ISSN
1680-0737
e-ISSN
—
Počet stran výsledku
7
Strana od-do
391-397
Název nakladatele
Springer Nature Singapore Pte Ltd.
Místo vydání
—
Místo konání akce
Prague
Datum konání akce
3. 6. 2018
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—