Evaluation of Electrical Impedance Tomography for a Measurement of Tidal Volume in High-frequency Ventilation of the Pig - Case Study
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F68407700%3A21460%2F19%3A00336873" target="_blank" >RIV/68407700:21460/19:00336873 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1109/EHB47216.2019.8970079" target="_blank" >http://dx.doi.org/10.1109/EHB47216.2019.8970079</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/EHB47216.2019.8970079" target="_blank" >10.1109/EHB47216.2019.8970079</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Evaluation of Electrical Impedance Tomography for a Measurement of Tidal Volume in High-frequency Ventilation of the Pig - Case Study
Popis výsledku v původním jazyce
We have analyzed dynamic lung impedance recorded by electric impedance tomography in the highfrequency oscillatory ventilation of the laboratory pig. The aim of the study was find out if electrical impedance tomography can be used for an assessing of the tidal volume in high-frequency oscillatory ventilation. The analyzed amplitude of dynamic lung impedance was compared with the tidal volume computed from flow that was measured at the orifice placed before endotracheal tube. We have found out that amplitude of dynamic lung impedance does not correspond with tidal volume within conducted experiment.
Název v anglickém jazyce
Evaluation of Electrical Impedance Tomography for a Measurement of Tidal Volume in High-frequency Ventilation of the Pig - Case Study
Popis výsledku anglicky
We have analyzed dynamic lung impedance recorded by electric impedance tomography in the highfrequency oscillatory ventilation of the laboratory pig. The aim of the study was find out if electrical impedance tomography can be used for an assessing of the tidal volume in high-frequency oscillatory ventilation. The analyzed amplitude of dynamic lung impedance was compared with the tidal volume computed from flow that was measured at the orifice placed before endotracheal tube. We have found out that amplitude of dynamic lung impedance does not correspond with tidal volume within conducted experiment.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20601 - Medical engineering
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
IEEE E-HEALTH AND BIOENGINEERING EHB 2019
ISBN
978-1-7281-2603-6
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
—
Název nakladatele
Gr. T. Popa University of Medicine and Pharmacy
Místo vydání
Iasi
Místo konání akce
Iasi
Datum konání akce
21. 11. 2019
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000558648300208